§ 175.69. Therapeutic Radiation Machines – Electronic Brachytherapy
(a) Applicability.
(1) An electronic brachytherapy device that does not meet the requirements of this section must not be used for irradiation of patients.
(2) An electronic brachytherapy device must only be utilized for human use applications that is specifically approved by the U.S. Food and Drug Administration, unless participating in a research study approved by the registrant's Institutional Review Board.
(3) Electronic brachytherapy devices are exempt from the requirements of 24 RCNY Health Code § 175.64.
(b) Possession of survey instruments. Each facility location authorized to use an electronic brachytherapy device in accordance with this section must possess appropriately calibrated portable monitoring equipment. At a minimum, such equipment must include a portable radiation measurement survey instrument capable of measuring dose rates over the range 10 µSv (1 mrem) per hour to 10 mSv (1,000 mrem) per hour. The survey instruments must be operable and calibrated in accordance with 24 RCNY Health Code § 175.66 for the applicable electronic brachytherapy source energy.
(c) Facility design requirements. In addition to shielding adequate to meet the requirements of 24 RCNY Health Code § 175.67, an electronic brachytherapy device treatment room must meet the following design requirements:
(1) If applicable, provision must be made to prevent simultaneous operation of more than one therapeutic radiation machine in a treatment room.
(2) Access to the treatment room must be controlled by a door at each entrance.
(3) Each treatment room must have provisions to permit continuous aural communication and visual observation of the patient from the treatment control panel during irradiation. The electronic brachytherapy device must not be used for patient irradiation unless the patient can be observed.
(4) For electronic brachytherapy devices capable of operating below 150 kV, radiation shielding for the staff in the treatment room must be available, either as a portable shield or as localized shielded material around the treatment site.
(5) For electronic brachytherapy devices capable of operating at greater than 150 kV:
(i) the control panel must be located outside the treatment room; and
(ii) electrical interlocks must be provided for all doors to the treatment room that will:
(A) prevent the operator from initiating the treatment cycle unless each treatment room entrance door is closed;
(B) cause the source to be shielded when an entrance door is opened; and
(C) prevent the source from being exposed following an interlock interruption until all treatment room entrance doors are closed and the source on-off control is reset at the console.
(d) Control panel. The control panel, in addition to the displays required by other provisions in this section, must:
(1) provide an indication of whether electrical power is available at the control panel and if activation of the electronic brachytherapy source is possible;
(2) provide an indication of whether x-rays are being produced;
(3) provide a means for indicating electronic brachytherapy source potential and current;
(4) provide the means for terminating an exposure at any time; and
(5) include an access control (locking) device that will prevent unauthorized use of the electronic brachytherapy device.
(e) Timer. A suitable irradiation control device (timer) must be provided to terminate the irradiation after a pre-set time interval or integrated charge on a dosimeter-based monitor.
(1) A timer must be provided at the treatment control panel. The timer must indicate planned setting and the time elapsed or remaining.
(2) The timer must not permit an exposure if set at zero.
(3) The timer must be a cumulative device that activates with an indication of "BEAM- ON" and retains its reading after irradiation is interrupted or terminated. After irradiation is terminated and before irradiation can be reinitiated, it must be necessary to reset the elapsed time indicator before irradiation can be resumed.
(4) The timer must terminate irradiation when a pre-selected time has elapsed, if any dose monitoring system has not previously terminated irradiation.
(5) The timer must permit setting of exposure times as short as 0.1 second.
(6) The timer must be accurate to within 1 percent of the selected value or 0.1 second, whichever is greater.
(f) QMP support.
(1) The services of a QMP must be required in facilities having electronic brachytherapy devices. The QMP must be responsible for:
(i) evaluation of the output from the electronic brachytherapy source;
(ii) generation of the necessary dosimetric information;
(iii) supervision and review of treatment calculations prior to initial treatment of any treatment site;
(iv) establishing the periodic and day-of-use quality assurance checks and reviewing the data from those checks as required by subdivision (j) of this section;
(v) consultation with the authorized user in treatment planning, as needed; and
(vi) performing calculations/assessments regarding patient treatments that may constitute a medical event.
(2) If the QMP is not a full-time employee of the registrant, the operating procedures required by subdivision (g) of this section must also specifically address how the QMP is to be contacted for problems or emergencies, as well as the specific actions, if any, to be taken until the QMP can be contacted.
(g) Operating procedures.
(1) Electronic brachytherapy devices must not be made available for medical use, unless the requirements of 24 RCNY Health Code § 175.61(a) and subdivisions (i) and (j) of this section have been met.
(2) The electronic brachytherapy device must be inoperable, either by hardware or password, when unattended by qualified staff or service personnel.
(3) Only individuals approved by the authorized user, radiation safety officer, or QMP shall be present in the treatment room during treatment.
(4) During operation, the electronic brachytherapy device operator must monitor the position of all persons in the treatment room, and all persons entering the treatment room, to prevent entering persons from unshielded exposure from the treatment beam.
(5) If a patient must be held in position during treatment, mechanical supporting or restraining devices must be used.
(6) Written procedures must be developed, implemented, and maintained for responding to equipment malfunction and any deviation from expected clinical outcomes. These procedures must include:
(i) instructions for responding to equipment failures and the names of the individuals responsible for implementing corrective actions; and
(ii) the names and telephone numbers of the authorized users, the QMP, and the radiation safety officer to be contacted if the device or console operates abnormally.
(7) A copy of the current operating and emergency procedures must be physically located at the electronic brachytherapy device control console. If the control console is integral to the electronic brachytherapy device, the required procedures must be kept where the operator is located during electronic brachytherapy device operation.
(8) Instructions must be posted at the electronic brachytherapy device control console to inform the operator of the names and telephone numbers of the authorized users, the QMP, and the radiation safety officer to be contacted if the device or console operates abnormally.
(9) The radiation safety officer or a designee, and an authorized user must be notified as soon as possible if a medical event occurs. The radiation safety officer or the QMP must inform the Department of the event.
(h) Safety precautions.
(1) A QMP must determine which persons in the treatment room require monitoring when the beam is energized.
(2) An authorized user and a QMP must be physically present during the initiation of all patient treatments involving the electronic brachytherapy device.
(3) A QMP and either an authorized user or a physician or electronic brachytherapy device operator, under the supervision of an authorized user, who has been trained in the operation and emergency response for the electronic brachytherapy device, must be physically present during continuation of all patient treatments involving the electronic brachytherapy device.
(4) When shielding is required by paragraph (4) of subdivision (c) of this section, the electronic brachytherapy device operator must use a survey meter to verify proper placement of the shielding immediately upon initiation of treatment. Alternatively, a QMP must designate shield locations sufficient to meet the requirements of this Article for any individual, other than the patient, in the treatment room.
(5) All personnel in the treatment room are required to remain behind shielding during treatment. A QMP must approve any deviation from this requirement and must designate alternative radiation safety protocols, compatible with patient safety, to provide an equivalent degree of protection.
(i) Source calibration measurements.
(1) Calibration of the electronic brachytherapy source output for an electronic brachytherapy device subject to this section must be performed by, or under the direct supervision of, a QMP.
(2) Calibration of the electronic brachytherapy source output must be made for each electronic brachytherapy source, or after any repair affecting the x-ray beam generation, or when indicated by the electronic brachytherapy source quality assurance checks.
(3) Calibration of the electronic brachytherapy source output must utilize a dosimetry system described in 24 RCNY Health Code § 175.61(c).
(4) Calibration of the electronic brachytherapy source output must include, as applicable, the determination of:
(i) the output within 2 percent of the expected value, if applicable, or determination of the output if there is no expected value;
(ii) timer accuracy and linearity over the typical range of use;
(iii) proper operation of back-up exposure control devices;
(iv) evaluation that the relative dose distribution about the source is within 5 percent of that expected; and
(v) source positioning accuracy to within 1 millimeter within the applicator;
(5) Calibration of the x-ray source output required by paragraphs (1) through (4) of this subdivision must be in accordance with current published recommendations from a recognized national professional association with expertise in electronic brachytherapy (when available). In the absence of a calibration protocol published by a national professional association, the manufacturer's calibration protocol must be followed.
(6) The registrant must maintain a record of each calibration in an auditable form for the duration of the registration. The record must include:
(i) the date of the calibration;
(ii) the manufacturer's name, model number and serial number for the electronic brachytherapy device and a unique identifier for its electronic brachytherapy source;
(iii) the model numbers and serial numbers of the instrument used to calibrate the electronic brachytherapy device; and
(iv) the name and signature of the QMP responsible for performing the calibration.
(j) Periodic and day-of-use quality assurance checks.
(1) Quality assurance checks must be performed on each electronic brachytherapy device subject to this section:
(i) at the beginning of each day of use;
(ii) each time the device is moved to a new room or site ("site" here is intended to include each day of use at each operating location for a self-contained electronic brachytherapy unit transported in a van or trailer); and
(iii) after each x-ray tube installation.
(2) The registrant must perform periodic quality assurance checks required by paragraph (1) of this subdivision in accordance with procedures established by the QMP.
(3) To satisfy the requirements of paragraph (1) of this subdivision, radiation output quality assurance checks must include as a minimum:
(i) verification that output of the electronic brachytherapy source falls within 3 percent of expected values, as appropriate for the device, as determined by:
(A) output as a function of time, or
(B) output as a function of setting on a monitor chamber.
(ii) verification of the consistency of the dose distribution to within 3 percent of that found during calibration required by subdivision (i) of this section.; and
(iii) validation of the operation of positioning methods to ensure that the treatment dose exposes the intended location within 1 mm; and
(4) The registrant must use a dosimetry system that has been intercompared within the previous 12 months with the dosimetry system described in 24 RCNY Health Code § 175.61(c)(1) to make the quality assurance checks required by paragraph (3) of this subdivision.
(5) The registrant must review the results of each radiation output quality assurance check according to the following procedures:
(i) an authorized user and QMP must be immediately notified if any parameter is not within its acceptable tolerance. The electronic brachytherapy device must not be made available for subsequent medical use until the QMP has determined that all parameters are within their acceptable tolerances
(ii) if all radiation output quality assurance check parameters appear to be within their acceptable range, the quality assurance check must be reviewed and signed by either the authorized user or QMP within 2 days; and
(iii) the QMP must review and sign the results of each radiation output quality assurance check at intervals not to exceed 30 days.
(6) To satisfy the requirements of paragraph (1) of this subdivision, safety device quality assurance checks must, at a minimum, assure:
(i) proper operation of radiation exposure indicator lights on the electronic brachytherapy device and on the control console;
(ii) proper operation of viewing and intercom systems in each electronic brachytherapy facility, if applicable;
(iii) proper operation of radiation monitors, if applicable;
(iv) the integrity of all cables, catheters or parts of the device that carry high voltages; and
(v) connecting guide tubes, transfer tubes, transfer-tube-applicator interfaces, and treatment spacers are free from any defects that interfere with proper operation.
(7) If the results of the safety device quality assurance checks required by paragraph (6) of this subdivision indicate the malfunction of any system, a registrant must secure the control console in the OFF position and not use the electronic brachytherapy device except as may be necessary to repair, replace, or check the malfunctioning system.
(8) The registrant must maintain a record of each quality assurance check required by paragraphs (3) through (7) of this subdivision in an auditable form for 3 years.
(i) The record must include the date of the quality assurance check; the manufacturer's name, model number and serial number for the electronic brachytherapy device; the name and signature of the individual who performed the periodic quality assurance check and the name and signature of the QMP who reviewed the quality assurance check;
(ii) For radiation output quality assurance checks required by paragraph (3) of this subdivision, the record must also include the unique identifier for the electronic brachytherapy source and the manufacturer's name; model number and serial number for the instruments used to measure the radiation output of the electronic brachytherapy device.
(k) Therapy-related computer systems. The registrant must perform acceptance testing on the treatment planning system of electronic brachytherapy-related computer systems in accordance with current published recommendations from a recognized national professional association with expertise in electronic brachytherapy (if available). In the absence of an acceptance testing protocol published by a national professional association, the manufacturer's acceptance testing protocol must be followed.
(1) Acceptance testing must be performed by, or under the direct supervision of, a QMP. At a minimum, the acceptance testing must include, as applicable, verification of:
(i) the source-specific input parameters required by the dose calculation algorithm;
(ii) the accuracy of dose, dwell time, and treatment time calculations at representative points;
(iii) the accuracy of isodose plots and graphic displays;
(iv) the accuracy of the software used to determine radiation source positions from radiographic images; and
(v) if the treatment-planning system is different from the treatment- delivery system, the accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.
(2) The position indicators in the applicator must be compared to the actual position of the source or planned dwell positions, as appropriate, at the time of commissioning.
(3) Prior to each patient treatment regimen, the parameters for the treatment must be evaluated and approved by the authorized user and the QMP for correctness through means independent of that used for the determination of the parameters.
(l) Training.
(1) A registrant must provide instruction, initially and at least annually, to all individuals who operate the electronic brachytherapy device, as appropriate to the individual's assigned duties, in the operating procedures identified in subdivision (g) of this section. If the interval between patients exceeds one year, retraining of the individuals shall be provided.
(2) In addition to the requirements of subdivisions (c) of 24 RCNY Health Code § 175.60 as to authorized users, these individuals must also receive device specific instruction initially from the manufacturer and annually from either the manufacturer or other qualified trainer. The training must be of a duration recommended by a recognized national professional association with expertise in electronic brachytherapy (if available). In the absence of any training protocol recommended by a national professional association, the manufacturer's training protocol must be followed. The training must include, but not be limited to:
(i) device-specific radiation safety requirements;
(ii) device operation;
(iii) clinical use for the types of use approved by the FDA;
(iv) emergency procedures, including an emergency drill; and
(v) the registrant's quality assurance program.
(3) A registrant must retain a record of individuals receiving instruction required by paragraphs (1) and (2) of this subdivision for 3 years. The record must include a list of the topics covered, the date of the instruction, the names of the attendees, and the names of the individuals who provided the instruction.
(m) Mobile electronic brachytherapy service. A registrant providing mobile electronic brachytherapy service must, at a minimum:
(1) check all survey instruments before medical use at each address of use or on each day of use, whichever is more restrictive;
(2) account for the electronic brachytherapy source in the electronic brachytherapy device before departure from the client's address;
(3) perform, at each location on each day of use, all of the required quality assurance checks specified in subdivision (j) of this section to assure proper operation of the device.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.70. Therapeutic Radiation Machines – Other Use of Electronically-Produced Radiation
A person must not utilize any device which is designed to electrically generate a source of ionizing radiation to deliver therapeutic radiation dosage, or which is not appropriately regulated under any existing category of therapeutic radiation machine, until:
(a) the applicant or registrant has, at a minimum, provided the Department with:
(1) a detailed description of the device and its intended application;
(2) facility design requirements, including shielding and access control;
(3) documentation of appropriate training for authorized user physicians and QMPs;
(4) methodology for measurement of dosages to be administered to patients or human research subjects;
(5) documentation regarding calibration, maintenance, and repair of the device, as well as instruments and equipment necessary for radiation safety;
(6) radiation safety precautions and instructions; and
(7) other information requested by the Department in its review of the application; and
(b) the applicant or registrant has received written approval from the Department to utilize the device in accordance with the requirements and specific conditions the Department considers necessary for the medical use of the device.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.100. Incorporation of Federal Regulations
(a) All persons subject to the requirements of this Part (24 RCNY Health Code §§ 175.100 through 175.108) are required to comply with the specific provisions of Title 10 of the Code of Federal Regulations ("C.F.R.") issued by the United States Nuclear Regulatory Commission ("NRC") as expressly indicated in this Part and which are hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth herein in their entirety. For the purposes of this Part, such incorporation by reference shall mean the specific provisions of Title 10 of the C.F.R. in effect as of the effective date of this Part, and any successor regulations promulgated by the NRC in Title 10 of the C.F.R. expressly indicated in this Part. Any such successor regulation shall be considered adopted in this Part as of the effective date of the provision in the C.F.R.
(b) In this Part (24 RCNY Health Code §§ 175.100 through 175.108), the following definitions apply:
(1) "10 C.F.R." or "C.F.R." means Title 10 of the Code of Federal Regulations, Chapter I (Nuclear Regulatory Commission).
(2) "City" means the 5 boroughs of New York City.
(3) Except as indicated in subdivision (c) of this section, references to "Department" means the New York City Department of Health and Mental Hygiene.
(4) "Radioactive material" means any solid, liquid, or gas which spontaneously emits alpha particles, beta particles, gamma rays, x rays, neutrons, high-speed electrons, high-speed protons, or other particles capable of producing ions. For the purposes of this Part, radioactive material does not include material that emits only non-ionizing radiation, such as radiowaves or microwaves, visible, infrared or ultraviolet light.
(c) To reconcile differences between this Article and the incorporated sections of the C.F.R., the following meanings are substituted for certain terms in the incorporated language of the C.F.R.:
(1) Except as indicated in paragraph (2) of this subdivision, any reference to ''NRC'' or ''Commission'' in the incorporated C.F.R. regulations means the Department. Any notifications and correspondence required to be sent to the NRC in the incorporated regulations of the C.F.R. should instead be sent to the Department at the address provided in 24 RCNY Health Code § 175.01(c).
(2) The following references to "NRC" or "Commission" in the incorporated C.F.R. regulations shall continue to refer to and mean the Nuclear Regulatory Commission:
(i) References in 10 C.F.R. § 30.12 to Commission contractors;
(ii) References in 10 C.F.R. Part 35 to Commission master material license or licensee;
(iii) References in 10 C.F.R. Part 35 to the "NRC's web page" for listing of acceptable board certifications;
(iv) References in 10 C.F.R. § 37.27 relating to criminal history record checks;
(v) Reference in 10 C.F.R. § 37.29(a)(1) to "an employee of the Commission";
(vi) References in 10 C.F.R. Part 71 to NRC-approved packaging.
(3) The references to "Department" in 10 C.F.R. §§ 30.12 and 30.14(b)(1) means the U.S. Department of Energy.
(4) Any reference to ''agreement state'' within the incorporated C.F.R. regulations means an external regulatory authority other than this Department.
(5) Any cross reference within the incorporated C.F.R. regulations means a cross reference to the C.F.R. unless otherwise specified;, for example, a reference to "§ 30.18 of this Chapter" means "10 C.F.R. § 30.18."
(6) References to forms in the incorporated C.F.R. regulations mean the appropriate forms prescribed by the Department.
(d) The C.F.R. provisions incorporated by reference herein may be obtained from:
(1) the Department at the address provided in 24 RCNY Health Code § 175.01(c), or
(2) the United States Government Publishing Office (GPO), 710 North Capitol Street, N.W., Washington, DC 20401 (866) 512-1800, or, online at (3) the U.S. Nuclear Regulatory Commission online at
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.101. Notices, Instructions and Reports to Workers
(a) Except as set forth in subdivision (b) of this section, 10 C.F.R. Part 19 is hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in its entirety.
(b) The following provisions from 10 C.F.R. Part 19 are not so incorporated: § 19.1, § 19.2, the definition of "regulated entities" in § 19.3, § 19.4, § 19.5, § 19.8, §§ 19.11(b) and (e), § 19.14(a), § 19.18, § 19.30, § 19.31, § 19.32 and§ 19.40.
(c) Additional requirements.
(1) Instructions to workers as required in 10 C.F.R. § 19.12 must be given to all individuals working in or frequenting any portion of a restricted area.
(2) In addition to the requirements of 10 C.F.R. § 19.12, individuals must be instructed in the operating procedures applicable to work under the license and must be required to demonstrate familiarity with precautions, procedures, and devices included in the instructions.
(3) Records documenting individual worker instruction shall be maintained for inspection by the Department for a period of 3 years.
(4) Each licensee must inform each worker annually of the worker's exposure to radiation or radioactive material as shown in the records maintained by the licensee pursuant to 10 C.F.R. § 20.2106.
(Added City Record 4/24/2019, eff. 5/24/2019; amended City Record 6/13/2025, eff. 7/13/2025)
§ 175.102. Standards for Protection Against Radiation
(a) Except as set forth in subdivision (b) of this section, 10 C.F.R. Part 20 is hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in its entirety.
(b) The following provisions from 10 C.F.R. Part 20 are not so incorporated: § 20.1001, § 20.1002, §§ 20.1006 through 20.1009, § 20.1406(b), § 20.1905(g), § 20.2203(c), § 20.2206, § 20.2401, § 20.2402 and Part 20 Appendix D.
(c) Additional requirements.
(1) Radiation protection programs.
(i) Each licensee must:
(A) provide a radiation safety officer pursuant to 10 C.F.R. §§ 35.24 and 35.50 who shall be delegated authority to ensure the implementation of this radiation protection program and ensure radiation doses are As Low As Reasonably Achievable (ALARA) (as defined in 10 C.F.R. § 20.1003 and more fully described in 24 RCNY Health Code § 175.105(c)(1)). The radiation safety officer or associate radiation safety officer named on the license, or an authorized user designated to act as the radiation safety officer in the radiation safety officer's absence, must be present on the premises at least 50 percent of the time that radioactive material is being handled or equipment containing radioactive material is being operated;
(B) provide for a radiation safety committee to administer the radiation protection program in medical centers, hospitals and institutions of higher education. The committee must include the facility operator or a person with the authority to act on behalf of the facility operator, and representation from departments within the facility where radiation sources are used. The committee must meet at least quarterly and must oversee all uses of radioactive materials within the facility, must review the activities of the radiation safety officer, and must review the radiation safety program at least annually. The committee, or a subcommittee, must oversee the administration of a quality assurance program as required by 24 RCNY Health Code § 175.108;
(C) provide a quality assurance program for diagnostic and therapeutic uses of radioactive materials pursuant to 24 RCNY Health Code § 175.108 and other applicable provisions of this Article; and
(D) ensure that all personnel involved in planning for, or administering radiation doses to humans, or in the use of radioactive materials for other lawful purposes, are supervised, are instructed as described in 10 C.F.R. § 19.12 and are competent to safely use such radiation sources and services.
(ii) For non-human use radioactive materials installations, the radiation safety officer specified in 24 RCNY Health Code § 175.102(c)(1)(i)(A) must be:
(A) a physicist certified by the American Board of Health Physics, the American Board of Radiology or the American Board of Medical Physics in a branch of physics related to the type and use of radioactive material in the installation; or
(B) a person with equivalent training and experience as determined by the Department; or
(C) an authorized user named on the radioactive materials license issued by the Department.
(2) Individual monitoring.
(i) In addition to the criteria in 10 C.F.R. § 20.1502(a)(2),(3), each licensee must monitor occupational exposure for minors and declared pregnant women likely to receive, in 1 year from sources external to the body, a deep dose equivalent in excess of 10 percent of any of the applicable limits in 10 C.F.R. §§ 20.1201 or 20.1208.
(ii) A person supplying personnel monitoring devices to individuals pursuant to 10 C.F.R. § 20.1502 must ensure that the individuals wear such devices as follows:
(A) An individual monitoring device used for monitoring the dose to the whole body must be worn at the unshielded location of the whole body likely to receive the highest exposure, except when monitoring is performed in accordance with 24 RCNY Health Code § 175.17(b)(4).
(B) An individual monitoring device used for monitoring the dose to an embryo/fetus of a declared pregnant worker pursuant to 10 C.F.R. § 20.1208 must be located at the waist under any protective garment worn by the declared pregnant worker.
(C) An individual monitoring device used for monitoring the lens dose equivalent must be located at the neck outside any protective garment worn by the individual, or at an unshielded location closer to the eye.
(D) An individual monitoring device used for monitoring the dose to the extremities must be worn on the extremity likely to receive the highest exposure. The device must be oriented to measure the highest dose to the extremity being monitored.
(iii) The licensee must ensure that adequate precautions are taken to prevent a deceptive exposure of an individual monitoring device, and no licensee must remove an exposure from an individual's exposure record without prior authorization from the Department. The licensee must submit the dosimeter for processing with due diligence and in no event in excess of the time period specified by the manufacturer of the dosimeter.
(3) Respiratory protection.
(i) If the licensee uses respiratory protection equipment to limit intakes pursuant to 10 C.F.R. §§ 20.1701 through 20.1705, the licensee must issue a written policy statement on respirator usage covering:
(A) the routine, non-routine, and emergency use of respirators; and
(B) limitations on periods of respirator use and relief from respirator use; and
(C) the use of process or other engineering controls, instead of respirators.
(ii) For a licensee to make allowance for respiratory equipment when estimating the exposure of individuals to airborne radioactive material, the following additional condition applies: the licensee selects respiratory protection equipment that provides a protection factor, specified in 10 C.F.R. Part 20 Appendix A, greater than the multiple by which peak concentrations of airborne radioactive materials in the working area are expected to exceed the values specified in Table 1, Column 3 of 10 C.F.R. Part 20 Appendix A. However, if the selection of respiratory protection equipment with a protection factor greater than this multiple of peak concentration is inconsistent with the goal specified in 10 C.F.R. § 20.1702 of keeping the total effective dose equivalent ALARA, the licensee may select respiratory protection equipment with a lower protection factor provided that such a selection would result in a total effective dose equivalent that is ALARA. In estimating the dose to individuals from intake of airborne radioactive materials, the concentration of radioactive material in the air that is inhaled when respirators are worn is initially assumed to be the ambient concentration in air without respiratory protection, divided by the assigned protection factor. If the dose is later found to be greater than the initially estimated dose, the corrected value must be used. If the dose is later found to be less than the initially estimated dose, the corrected value may be used.
(4) Requirements for possession of sealed sources.
(i) The licensee in possession of any sealed source must follow the radiation safety and leak testing requirements specified in 10 C.F.R. § 35.67.
(ii) Reports of test results for leaking or contaminated sealed sources must be made pursuant to 10 C.F.R. § 35.3067.
(iii) A licensee authorized for medical use in possession of a licensed sealed source or brachytherapy source must survey with a radiation survey instrument at intervals not to exceed 3 months all areas where such sources are stored. This shall not apply to sources in teletherapy units or gamma stereotactic radiosurgery units or to sealed sources in diagnostic devices.
(5) Radioactive material must not be stored with either food or beverages.
(6) No person shall bury any licensed radioactive material within New York City.
(7) The provisions of 10 C.F.R. § 20.2005 do not authorize the licensed materials described to be disposed of inside the City as if they were not radioactive; however, these materials may be shipped for disposal outside of the City as if they were not radioactive provided that the receiving jurisdiction allows the disposal of such materials as if they were not radioactive.
(8) Vacating premises. Each specific licensee must notify the Department in writing of intent to vacate not less than 30 days before vacating or relinquishing possession or control of premises which may have been contaminated with radioactive material as a result of licensed activities. When deemed necessary by the Department, the licensee must decontaminate the premises to such levels as the Department may specify.
(9) Recordkeeping requirements for receipt, use, and disposition of radioactive material.
(i) Each licensee must maintain records of the receipt, use, and disposition of radioactive material in units of becquerels or microcuries and must include from whom such materials were received and the ultimate disposition.
(ii) The licensee must retain these records for 3 years after the record is made.
(10) The Department may order the removal, through an authorized person, or the surrender to the Department, of any radiation source by any person who:
(i) does not hold, or continue to hold, a valid license issued by the Department; or
(ii) is not able or equipped, or who fails to observe with regard to such radiation source, those radiation protection standards established or enforced by the Department or who uses such radiation source in violation of law, regulation, this Article, order or license issued by the Department. Such person must decontaminate any premises which may have been contaminated with radioactive material as a result of any such activities to such radiation levels as the Department may specify. The expenses incidental to such transfer, surrender, and decontamination must be borne by such person responsible for the source.
(11) In determining the extent of any individual's exposure to radiation subsequent to any radiation accident, contamination, theft or loss, the licensee must comply with all orders of the Department directing such licensee to make available to such individual appropriate medical evaluation services or appropriate tests and to furnish a copy of the reports of such evaluation or test to the Department.
(12) A licensee or applicant for a license must obtain any permits required by the New York State Department of Environmental Conservation pursuant to 6 NYCRR Part 380, or any successor law or regulation, and must develop, document, and implement a discharge minimization program required by the New York State Department of Environmental Conservation pursuant to 6 NYCRR § 380-7, or any successor law or regulation.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.103. General Requirements for Radioactive Materials
(a) Except for the removal of source material from its place of origin in nature, or as otherwise may be provided in this Article, no person subject to this Article shall transfer, receive, produce, possess or use in New York City any radioactive material except pursuant to a license issued by the Department.
(b) Except as set forth in subdivision (c) of this section, 10 C.F.R. Part 30 and the following provisions from 10 C.F.R. Part 40: § 40.3, the definition of "byproduct material and "depleted uranium" in § 40.4 § 40.11, § 40.12(a), § 40.13 (except § 40.13(c)(5)(iv)), § 40.21, § 40.22, § 40.41(a)-(c),(f), §§ 40.42 (c)-(k)(3),(l), § 40.46, § 40.51 (except 40.51(b)(6)), § 40.54, § 40.55, § 40.60 and § 40.61 are hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in their entirety.
(c) The following provisions from 10 C.F.R. Parts 30 are not so incorporated: § 30.1, § 30.2, the definitions of "commencement of construction" and "construction" in § 30.4, §§ 30.5 through 30.8, § 30.21(c), § 30.32(e), § 30.34(d), § 30.34(e)(1), § 30.34(e)(3), §§ 30.37 through 30.39, § 30.41(b)(6), § 30.53, 30.55, § 30.62, § 30.63 and § 30.64.
(d) References to "Department" in the incorporated regulations of 10 C.F.R. §§ 30.12 and 30.41(b)(1) means the U.S. Department of Energy.
(e) Any reference to "byproduct material" in the incorporated regulations of 10 C.F.R. §§ 30.31 through 30.62 means "radioactive material" as defined in this Article.
(f) Additional requirements.
(1) License termination requirements.
(i) If a licensee does not submit an application for renewal of a radioactive materials license pursuant to 10 C.F.R. § 30.32, then the licensee must, on or before the expiration date stated in the license,:
(A) Terminate use of radioactive material;
(B) Dispose of all radioactive material in accordance with all applicable regulations in effect at the time of disposal;
(C) Submit a written certification of the disposition of all radioactive materials authorized by the license on forms prescribed by the Department;
(D) Remove radioactive contamination to the extent practicable; and
(E) Conduct a radiation survey of the premises where the licensed activities were carried out and submit a copy of this survey to the Department. Such survey must be subject to approval by the Department and must include:
a. Levels of radiation in units (or multiples) of Gy-hr-1 (millirads-hr-1) at 1 cm for beta-gamma radiation or at 1 m for gamma radiation;
b. Levels of removable and fixed contamination, including alpha, in units of disintegrations (transformations) per min (becquerels) per 100 cm 2 for surfaces;
c. becquerels-ml-1 (mCi-ml-1) for water;
d. becquerels-kg-1 (pCi-g-1) for solids such as soil or concrete; and
e. a description of the survey or other measuring instruments used, including manufacturer and model numbers and date of most recent calibration.
(F) If the information submitted pursuant to clauses (A) through (E) of subparagraph (i) of paragraph (1) of subdivision (f) of this section does not adequately demonstrate that the premises are suitable for unrestricted use, the Department shall inform the licensee of the appropriate further actions required for the termination of the license, including, but not limited to, decontamination of the licensed premises to such levels and within such time frames as the Department may prescribe.
(ii) Each specific license shall continue in effect, beyond the expiration date if necessary, with respect to the presence of any residual radioactive materials that may be present as contamination until the Department terminates the license. During this time, the licensee must:
(A) Limit activities involving radioactive material to those related to decommissioning; and
(B) Continue to control entry to restricted areas until the Department determines they are suitable for release for unrestricted use and the Department terminates the license.
(iii) The Department can terminate a specific license when the Department determines that:
(A) All licensed radioactive material has been properly transferred or disposed; and
(B) premises have been decontaminated to such levels that the total effective dose equivalent (TEDE) from residual radioactivity distinguishable from background radiation, to an average member of the public will not exceed 25 mrem (0.25 mSv) per year;
(C) a radiation survey has been performed which describes all radiation levels and levels of fixed and removable contamination; and
(D) the licensee submits sufficient documentation to support a determination that the requirements of clauses (A) through (E) of subparagraph (i) of paragraph (1) of subdivision (f) of this section have been met.
(2) Amendment of licenses by the Department.
(i) A corrective amendment of any license may be issued by the Department at any time upon its initiative.
(ii) Any license may be amended or revoked by the Department by reason of the amendment of this Article, or any other applicable law.
(3) The licensee must notify the Department, in writing, within 30 days if an authorized user, radiation safety officer or radiation therapy physicist permanently discontinues performance of duties under the license.
(4) No person, in any advertisement or public posting, expressly or by implication, shall refer to the fact that a radiation installation is licensed by the Department, and no person shall state or imply that a radiation installation or its activities have been approved by the Department, the Board of Health or the Commissioner.
(5) Reciprocity. The holder of a license issued by the New York State Department of Environmental Conservation, the New York State Department of Health, the U.S. Nuclear Regulatory Commission or any Agreement State, may bring, possess or use radioactive material covered by such license within the Department's jurisdiction for a period not in excess of 30 days in any 12 consecutive months without obtaining a license from the Department, provided that:
(i) such license does not limit the holder's possession or use of such material to a specific installation or installations;
(ii) such holder, prior to bringing such material into the City, files with the Department a notice indicating the period, type and location of proposed possession and use within the Department's jurisdiction, and a copy of the license;
(iii) such holder supplies such additional information as the Department may reasonably request;
(iv) such holder, during the period of this possession and use of such material within the City, complies with all applicable sections of this Article except 24 RCNY Health Code § 175.103(a); and
(v) such holder, during such period, complies with all the terms and conditions of his license, except if any such license terms or conditions are determined by the Department to be inconsistent with the requirements of this Article.
(6) If the holder of a license issued by this Department, the New York State Department of Environmental Conservation, the New York State Department of Health, or an Agreement State intends to conduct any licensed activity in areas of exclusive federal jurisdiction within New York City, then before engaging in any such licensed activity for the first time in a calendar year, such licensee must provide the U.S. Nuclear Regulatory Commission with at least 3 days advanced notice of its proposed activity in such areas under exclusive federal jurisdiction within New York City.
(Added City Record 4/24/2019, eff. 5/24/2019; amended City Record 6/13/2025, eff. 7/13/2025)
§ 175.104. Specific Types of Radioactive Materials Licenses
(a) Types of license. The requirements specified in this section are in addition to, and not in substitution for, others in this Article. In particular, the incorporated provisions of 10 C.F.R. Part 30 apply to all license applications and all specific radioactive materials licenses. For the purposes of this Article, the following specific license types of radioactive materials apply:
(1) A specific license of limited scope for gamma stereotactic radiosurgery (GSR) or for teletherapy means a license that authorizes receipt, acquisition, ownership, possession, use and transfer of specified quantities and types of radioactive material for use in GSR or teletherapy programs.
(2) A specific license of limited scope for medical use means a license that authorizes receipt, acquisition, ownership, possession, use and transfer of specified quantities and types of radioactive material for use in or on humans in a medical program, but does not include GSR or teletherapy.
(3) A specific license of limited scope for other use means a license that authorizes receipt, production, acquisition, ownership, possession, use and transfer of specified quantities and types of radioactive material for uses other than in or on humans.
(4) A specific license of broad scope for medical use means a license that authorizes receipt, acquisition, ownership, possession, use and transfer of any chemical or physical form of the radioactive materials specified in the license, for use in or on humans in a medical program, but does not include GSR or teletherapy.
(5) A specific license of broad scope for research and development means a license that authorizes receipt, acquisition, ownership, possession, use and transfer of any chemical or physical form of radioactive materials specified in the license, in quantities not exceeding those specified in the license, for uses other than in or on humans.
(b) Specific licenses for human use of radioactive materials in institutions. An application by a medical institution for a specific license for medical use of radioactive material may be approved if:
(1) pursuant to §§ 175.103 and 175.105, the applicant satisfies the requirements specified in the incorporated provisions of 10 C.F.R. Parts 30 and 35; and
(2) the applicant possesses adequate facilities for the clinical care of patients; and
(3) the physician designated on the application as the individual authorized user has training and experience as specified in 10 C.F.R. Part 35, in the proposed use, handling and administration of radionuclides and the clinical management of radioactive patients. The physician must furnish evidence of such experience with the application. A statement from the physician's preceptor at the medical institution where the physician acquired such training and experience, indicating their amount and nature, may be submitted as evidence of such experience.
(4) The license application is signed by the chairman of the radiation safety committee and an authorized representative of the medical institution.
(c) Specific licenses to individual physicians for human use of radioactive materials. An application by an individual physician for a specific license for human use of radioactive material may be approved if:
(1) pursuant to §§ 175.103 and 175.105, the applicant satisfies the requirements specified in the incorporated provisions of 10 C.F.R. Parts 30 and 35; and
(2) the applicant has training and experience as specified in 10 C.F.R. Part 35 in the proposed use, handling and administration of radionuclides, and the clinical management of radioactive patients. The physician must furnish evidence of such training and experience with the application. A statement from the physician's preceptor at the institution where the physician acquired such training and experience, indicating their amount and nature, may be submitted as evidence of such experience.
(d) Specific licenses of broad scope.
(1) A specific license of broad scope is a specific license authorizing receipt, acquisition, ownership, possession, use, and transfer of any chemical or physical form of the byproduct material specified in the license, but not exceeding quantities specified in the license. Approved authorized users do not need to be listed on the license. Specific licenses of broad scope for medical use are exempted from certain requirements as listed in 10 C.F.R. § 35.15.
(2) A specific license of broad scope shall be issued only to medical institutions, research laboratories, or institutions of higher education; such licenses shall not be issued to individuals.
(3) An application for a specific license of broad scope may be approved if:
(i) the applicant satisfies the general requirements specified in 10 C.F.R. § 30.33; and
(ii) the applicant has engaged in a reasonable number of activities involving the use of radioactive material as determined by the Department; and
(iii) the applicant has established administrative controls and provisions relating to organization and management, procedures, recordkeeping, material control, and accounting and management review that are necessary to assure safe operations, including:
(A) the establishment of a radiation safety committee pursuant to 24 RCNY Health Code § 175.102(c)(1)(i)(B); and
(B) the appointment of a full-time radiation safety officer pursuant to 10 C.F.R. Part 20; and
(C) the establishment of appropriate administrative procedures to assure:
a. control of procurement and use of radioactive material; and
b. completion of safety evaluations of proposed uses of radioactive material which take into consideration such matters as the adequacy of facilities and equipment, training and experience of the user, and the operating or handling procedures; and
c. review, approval, and recording by the radiation safety committee of safety evaluations of proposed uses prepared in accordance with 24 RCNY Health Code § 175.104(d)(3)(iii)(C)(b) prior to the use of radioactive materials.
(4) The following are conditions of all specific licenses of broad scope. Unless specifically authorized pursuant to other provisions of this Article, broad scope licensees must not:
(i) conduct tracer studies in the environment involving direct release of radioactive material;
(ii) receive, acquire, own, possess, use, transfer, or import devices containing 3.7 E6 GBq (100,000 Ci) or more of radioactive material in sealed sources used for irradiation of materials;
(iii) add, or cause the addition of, radioactive material to any food, beverage, cosmetic, drug or other product designed for ingestion by, or application to, a human being except as authorized in the license.
(e) Specific licenses for non-human use. An application for a specific license authorizing non-human use of radioactive materials may be approved if:
(1) the applicant satisfies the general requirements specified in 10 C.F.R. § 30.33; and
(2) the applicant, or the applicant's personnel, has training and experience commensurate with the proposed amounts, types and uses of radioactive materials which must minimally include:
(i) a college degree at the bachelor level in a physical, biological, environmental or engineering science; and
(ii) at least 40 hours of training and experience in the safe handling of radioactive materials appropriate to the type and forms of such materials to be used, which must include:
(A) characteristics of ionizing radiation;
(B) units of radiation dose and quantities;
(C) radiation detection instrumentation; and
(D) biological hazards of exposure to radiation.
(f) General licenses.
(1) Except as set forth in paragraph (2) of this subdivision, 10 C.F.R. Part 31 is hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in its entirety.
(2) The following provisions from 10 C.F.R. Part 31 are not incorporated: § 31.1, § 31.2, § 31.4, § 31.22 and § 31.23.
(3) Any reference to "byproduct material" in the incorporated 10 C.F.R. § 31.9 means "radioactive material" as defined in this Article.
(4) Reference to "non-Agreement State" in the incorporated 10 C.F.R. § 31.6 means this Department. (5) Licensees exempt pursuant to incorporated 10 C.F.R. § 31.12(b) are similarly exempt from the analogous license requirements of this Article.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.105. Medical Use of Radioactive Materials
(a) Except as set forth in subdivision (b) of this section, 10 C.F.R. Part 35 is hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in its entirety.
(b) The following provisions from 10 C.F.R. Part 35 are not so incorporated: § 35.1, § 35.8, § 35.11(c)(1), § 35.13(a)(1), § 35.4001 and § 35.4002.
(c) Additional requirements.
(1) ALARA program.
(i) Each licensee must develop and implement a written program to maintain radiation doses and releases of radioactive material in effluents to unrestricted areas to be As Low As Reasonably Achievable (ALARA) in accordance with this subdivision.
(ii) To satisfy the requirement of subparagraph (i) of this paragraph (1):
(A) for licensees that are medical institutions, the management, radiation safety officer and all authorized users must participate in the establishment, implementation, and operation of the program as required by this Article or required by the radiation safety committee; or
(B) for licensees that are not medical institutions, management and all authorized users must participate in the program as requested by the radiation safety officer.
(iii) The ALARA program must include notice to workers of the program's existence and workers' responsibility to help keep dose equivalents ALARA.
(iv) For licensees that are medical institutions, the ALARA program must include an annual review by the radiation safety committee as indicated in paragraph (1) of subdivision (c) of 24 RCNY Health Code § 175.102.
(v) For licensees that are not medical institutions, the ALARA program must include an annual management review of summaries of the types and amounts of radioactive material used, occupational dose reports, and continuing education and training for all personnel who work with or in the vicinity of radioactive material, prepared by all authorized users or the radiation safety officer.
(vi) The purpose of the review required by subparagraphs (iv) and (v) of this paragraph (1) is to ensure that every reasonable effort to maintain occupational doses, doses to the general public, and releases of radioactive material to unrestricted areas, are ALARA, taking into account the state of technology and the cost of improvements in relation to benefits.
(vii) The licensee must retain a current written description of their ALARA program for the duration of the license. The written description must include:
(A) a commitment by management to keep occupational doses as low as reasonably achievable;
(B) a requirement that the radiation safety officer brief management at least once each year on the radiation safety program; personnel exposure investigational levels that, when exceeded, will initiate an investigation by the radiation safety officer of the cause of the exposure (ALARA I); and personnel exposure investigational levels that, when exceeded, will initiate a prompt investigation by the radiation safety officer of the cause of the exposure and a consideration of actions that might to be taken to reduce the probability of recurrence (ALARA II).
(2) Qualified personnel. Personnel, duly licensed by the New York State Department of Health to practice nuclear medicine technology (other than physicians or registered professional nurses), at licensee locations involved in the performance of diagnostic procedures utilizing radioactive material, which includes parenteral administration of radioactive material by intravenous, intramuscular or subcutaneous methods, must:
(i) have satisfactorily completed an educational program in nuclear medicine technology accredited by the Committee on Allied Health Education and Accreditation, or the accrediting agency of the state in which the program was completed, provided such state accreditation requires education and training in the above methods of parenteral administration; or
(ii) possess certification as a nuclear medicine technologist by the American Registry of Radiologic Technologists or certification by the Nuclear Medicine Technology Board; and
(iii) prior to permitting parenteral administration by a nuclear medicine technologist, the medical board of a hospital, a physician, or the radiation safety committee of an institution who have no medical board, must adopt with governing authority approval:
(A) procedures to assure that the nuclear medicine technologist possesses the education and training or certification set forth in this subdivision and is proficient in the competent performance of parenteral administration; and
(B) requirements for authorized user physician, which at a minimum must require supervision by such physician when parenteral administration of radioactive material for diagnostic testing is performed by a qualified nuclear medicine technologist.
(iv) The requirements of this paragraph concerning qualified personnel are in addition to, and not in place of, any other requirements in this Article or in the incorporated 10 C.F.R. sections relating to supervision or the presence of an authorized user.
(3) Possession, use, calibration and check of dose calibrators.
(i) A medical use licensee authorized to administer radioactive materials must possess a dose calibrator and use it to measure the amount of activity administered to each patient.
(ii) A licensee must:
(A) check each dose calibrator for constancy with a dedicated check source at the beginning of each day of use. To satisfy this requirement, the check must be done on a frequently used setting with a sealed source of any photon-emitting radionuclide with a half-life greater than 90 days;
(B) test each dose calibrator for accuracy upon installation and at intervals not to exceed 12 months thereafter by assaying at least 2 sealed sources of appropriate source activity, containing different radionuclides, the activity of which the manufacturer has determined by traceability to a national standard to be within 5 percent of the stated activity;
(C) test each dose calibrator for linearity upon installation and at intervals not to exceed 3 months thereafter over the range of use between 1 MBq (27 µCi) and the highest dosage that will be administered; and
(D) test each dose calibrator for geometry dependence upon installation over the range of volumes and volume configurations for which it will be used. The licensee must keep a record of this test for the duration of the use of the dose calibrator.
(iii) Notwithstanding the provisions of subparagraph (ii) of paragraph (3) of this subdivision, a licensee that uses a dose calibrator to measure the activity of beta- emitting radioactive materials to be administered to a patient must perform additional checks specified in clauses (A) and (B) of subparagraph (ii) of paragraph (3) of this subdivision using the same radionuclide to be administered and having an activity of at least 50 percent, but not more than 200 percent, of the prescribed activity or by equivalent procedures approved by the Department. Records must be kept pursuant to 24 RCNY Health Code § 175.105(c)(3)(vi).
(iv) A licensee must mathematically correct dosage readings for any geometry or linearity error that exceeds ±10 percent if the dosage is greater than 370 kBq (10µCi) and must repair or replace the dose calibrator if the accuracy or constancy error exceeds ±10 percent.
(v) A licensee must also perform checks and tests required by 24 RCNY Health Code § 175.103(c)(3)(ii) following adjustment or repair of the dose calibrator.
(vi) A licensee must retain a record of each check and test required by subparagraphs (ii), (iii) and (v) of 24 RCNY Health Code § 175.105(c)(3) for 3 years. Such records must include:
(A) for clause (A) of 24 RCNY Health Code § 175.105(c)(3)(ii), the models and serial numbers of the dose calibrator and check source, the identity and calibrated activity of the radionuclide contained in the check source, the date of the check, the activity measured, the instrument settings, and the name of the individual who performed the check;
(B) for clause (B) of 24 RCNY Health Code § 175.105(c)(3)(ii), the model and serial number of the dose calibrator, the model and serial number of each source used and the identity of the radionuclide contained in the source and its activity, proof of traceability to a national standard, the date of the test, the results of the test, the instrument settings, and the signature of the radiation safety officer; (C) for clause (C) of 24 RCNY Health Code § 175.105(c)(3)(ii), the model and serial number of the dose calibrator, the calculated activities, the measured activities, the date of the test, and the signature of the radiation safety officer; and
(D) for clause (D) of 24 RCNY Health Code § 175.105(c)(3)(ii), the model and serial number of the dose calibrator, the configuration and calibrated activity of the source measured, the activity of the source, the activity measured and the instrument setting for each volume measured, the date of the test, and the signature of the radiation safety officer.
(4) Possession and calibration of portable survey instruments.
(i) A licensee authorized to use radioactive material for any medical use permitted in 10 C.F.R. §§ 35.100, 35.200, 35.300, 35.400, 35.500, or 35.600 must have in its possession a portable, radiation detection survey instrument capable of detecting dose rates over the range of 1.0 µSv (0.1 mrem) per hour to 1,000 µSv (100 mrem) per hour, and a portable radiation measurement survey instrument capable of measuring dose rates over the range 10 µSv (1 mrem) per hour to 10 mSv (1,000 mrem) per hour. This requirement may be met by possession of two survey instruments or by possession of one survey instrument that meets both criteria. The instruments must be operable and calibrated in accordance with 10 C.F.R. § 35.61.
(ii) To meet the requirements of 24 RCNY Health Code § 175.105(c)(4)(i), the licensee must perform such calibrations as authorized by specific license condition, or must obtain the services of persons licensed by the U.S. Nuclear Regulatory Commission or an agreement state to perform calibrations of survey instruments.
(iii) A licensee must check each survey instrument for proper operation with the dedicated check source before each use. The licensee is not required to keep records of these checks.
(5) Surveys of exposure rate and contamination. In addition to the requirements of 10 C.F.R. § 35.70, a licensee must:
(i) survey with a radiation detection survey instrument at least once each week all areas where unsealed byproduct materials or radioactive wastes are stored;
(ii) conduct the surveys required by 10 C.F.R. § 35.70 and 24 RCNY Health Code § 175.105(c)(5)(i) so as to be able to detect and measure dose rates as low as 1 µSv (0.1 mrem) per hour.
(iii) establish dose rate action levels for the surveys required by 10 C.F.R. § 35.70 and 24 RCNY Health Code § 175.103(c)(5)(i) and must require that the individual performing the survey immediately notify the radiation safety officer if a dose rate exceeds an action level.
(iv) perform wipe tests for removable contamination once each week on all areas where radioactive materials are routinely prepared for use or administered and where unsealed sources of radioactive materials are stored, so as to be able to detect contamination on each wipe sample of 35 Bq (2100 disintegrations or transformations per minute).
(v) establish removable contamination action levels for the surveys required by 24 RCNY Health Code § 175.105(c)(5)(iv) and must require that the individual performing the survey immediately notify the radiation safety officer if contamination exceeds action levels.
(vi) retain a record of each survey or wipe test required by 10 C.F.R. § 35.70 and 24 RCNY Health Code § 175.105(c)(5) for 3 years. The record must include the date of the survey, a sketch of each area surveyed, action levels established for each area, the measured dose rate at several points in each area expressed in Sv (mrem) per hour or the removable contamination in each area expressed in becquerels (disintegrations or transformations per minute) per 100 square centimeters, the serial number and the model number of the instrument used to make the survey or analyze the samples, and the initials of the individual who performed the survey.
(6) Additional requirements for use of generators aerosols, gases, and volatile materials.
(i) A licensee approved to use unsealed byproduct material for medical use permitted in 10 C.F.R. § 35 Subpart D may use generators, radioactive aerosols, or radioactive gases only if specific application is made to and approved by the Department. (ii) Storage of volatiles and gases.
(A) A licensee must store volatile radiopharmaceuticals and radioactive gases in the shippers' radiation shield and container.
(B) After drawing the first dosage, a licensee must store and use a multidose container in a properly functioning fume hood.
(iii) Control of aerosols and gases.
(A) A licensee who administers radioactive aerosols or gases must do so with a system that will keep airborne concentrations within the limits prescribed by 10 C.F.R. Part 20.
(B) The system must provide for collection and decay or disposal of the unused aerosol or gas in a shielded container.
(C) Before receiving, producing, using, or storing a radioactive gas, the licensee must calculate the amount of time needed after a release to reduce the concentration in the area of use to the ALI listed in Table 1 of Appendix A of 10 C.F.R. Part 20. The calculation must be based on the highest activity of gas handled in a single container and the measured available air exhaust rate.
(D) A licensee must post the time calculated in 24 RCNY Health Code § 175.105(c)(6)(iii)(C) at the area of use, as well as safety measures to be instituted in case of a spill at the area of use.
(E) A licensee must check the operation of collection systems monthly and measure the ventilation rates in areas of use at intervals not to exceed 6 months. Records of these checks and measurements must be maintained for 3 years.
(F) A copy of the calculations, including assumptions, measurements and calculations made, required in 24 RCNY Health Code § 175.105(c)(6)(iii)(C) must be recorded and retained for the duration of the license.
(7) Radioactive cadavers.
(i) If any patient containing radioactive material administered/implanted for therapeutic purposes dies, the physician who pronounces such patient as dead must notify immediately the physician in charge of the case or such physician's designated representative.
(ii) No person shall commence any autopsy on any cadaver that contains more than 185 MBq (5 mCi) of radioactive material administered/implanted for therapeutic purposes without first having consulted with, and having been advised by, the radiation safety officer of the hospital or the physician responsible for the administration/implantation of the radioactive material. If neither is available, a designated representative may serve.
(iii) A radioactivity report on every cadaver containing more than 185 MBq (5 mCi) of radioactive material administered/implanted for therapeutic purposes must be completed by the radiation safety officer or the physician responsible for the administration of the radioactive material or their designated representative. The report must include the name, address and radioactive materials license number of the hospital; the name of the deceased; the name, address and telephone number of the next of kin; the name, address and telephone number of the funeral home to which the deceased will be sent; the radionuclide involved; the approximate activity on the date of the report and the physical form; the location of the radioactive materials within the body and the external dose rate at the body surface closest to the source; the precautions to be observed during autopsy or handling of the body by the funeral director; and the name of the person who prepared the form. This report must accompany the body, whether autopsied or not, when it is surrendered to the funeral director. The Department must be notified in person, by telephone, or by facsimile within 24 hours of the death and a copy of the radioactivity report must be sent to the Department within 15 days of the date of death.
(8) Safety requirements for uses authorized in 10 C.F.R. § 35 Subparts E and F. For each patient or research subject who cannot be released pursuant to 10 C.F.R. § 35.75, a licensee must:
(i) authorize visits by individuals under age 18 only on a patient-by- patient basis with the approval of the authorized user after consultation with the radiation safety officer.
(ii) promptly after implanting the dosage or sources, measure the dose rates in contiguous restricted and unrestricted areas with a radiation measurement survey instrument to demonstrate compliance with the requirements of 10 C.F.R. Part 20 and retain for 3 years a record of each survey that includes the time and date of the survey, a sketch of the area or list of points surveyed, the measured dose rate at several points expressed in mSv (mrem) per hour, the instrument used to make the survey, and the initials of the individual who made the survey.
(9) Additional requirements for uses authorized in 10 C.F.R. § 35 Subpart H.
(i) Amendment requests. In addition to the requirements specified in 10 C.F.R. § 35.13, a licensee must apply for and have received a license amendment prior to:
(A) making any change in the treatment room shielding;
(B) making any change in the location of a gamma stereotactic radiosurgery (GSR) unit or a teletherapy unit within the treatment room;
(C) using the GSR or teletherapy unit in a manner that could result in increased radiation levels in areas outside the treatment room;
(D) relocating the GSR or teletherapy unit; or
(E) allowing an individual not listed on the licensee's license to perform the duties of the authorized medical physicist.
(ii) A licensee must furnish a copy of the records required in 10 C.F.R. §§ 35.632 and 35.635 and the output from the teletherapy source expressed as Sv (rem) per hour at one meter from the source determined during the surveys required in 10 C.F.R. § 35.652 to the Department within 30 days following completion of the action that initiated the record requirement.
(iii) Modification of a GSR or teletherapy unit or room before beginning a treatment program. If the survey required by 10 C.F.R. § 35.652 or 24 RCNY Health Code § 175.105(c)(9)(iv) indicates that an individual in an unrestricted area may be exposed to levels of radiation greater than those permitted by 10 C.F.R. Part 20, before beginning the treatment program, the licensee must:
(A) either equip the unit with stops or add additional radiation shielding to ensure compliance with 10 C.F.R. Part 20;
(B) perform the survey required by 10 C.F.R. § 35.652 again; and
(C) include in the report required by 24 RCNY Health Code § 175.105(c)(9)(ii), the results of the initial survey, a description of the modification made to comply with 10 C.F.R. § 35.652(a) and the results of the second survey; or
(D) request and receive a license amendment that authorizes radiation levels in unrestricted areas greater than those permitted by 10 C.F.R. Part 20.
(10) Records.
(i) In lieu of the retention requirements of 10 C.F.R. § 35.2067, records of leak tests required by 10 C.F.R. § 35.67(b) and the semi-annual physical inventory of sealed sources and brachytherapy sources required by 10 C.F.R. § 35.67(g) shall be retained for 5 years.
(ii) In addition to the requirements of 10 C.F.R. § 35.2630, for each intercomparison, the record must include evidence that the intercomparison reading was sanctioned by a calibration laboratory or radiologic physics center accredited by the American Association of Physicists in Medicine.
(11) Visiting authorized user. Notwithstanding other provisions of this Article, a licensee may permit any physician to act as a visiting authorized user under the term of the license for up to 60 days per calendar year if:
(i) the visiting authorized user has the prior written permission of the licensee's management and, if the use occurs on behalf of an institution, the institution's Radiation Safety Committee (where applicable); and
(ii) the visiting authorized is specifically named as a user on a license issued by the Department, Nuclear Regulatory Commission, or an Agreement State, and only performs procedures for which they are specifically authorized on that license, and
(iii) the licensee maintains copies of the written permission required by subparagraph (i) of this paragraph and documentation that the visiting authorized user met the requirements of subparagraph (ii) of this paragraph for 5 years from the date of the last visit.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.106. Transportation and Packaging of Radioactive Materials
(a) Except as set forth in subdivision (b) of this section, 10 C.F.R. Part 71 is hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in its entirety.
(b) The following provisions from 10 C.F.R. Part 71 are not so incorporated: § 71.2, § 71.6, § 71.11, § 71.14(b), § 71.19, §§ 71.31 through 71.45, §§ 71.51 through 71.77, § 71.85(a)- (c), § 71.91(b), § 71.99, § 71.100, § 71.101(c)(2), § 71.101(d)-(f), § 71.107 through § 71.125.
(c) Licensees must ascertain that the certificate holder determinations in subsections (a) through (c) of 10 C.F.R. § 71.85 have been made.
(d) The requirements of the incorporated 10 C.F.R. §§ 71.106 and 71.135 apply to general licensees and do not apply to certificate of compliance holders or applicants for certificates of compliance as those terms are defined in 10 C.F.R. § 71.4.
(e) A licensee's quality assurance program referenced in the incorporated 10 C.F.R. Part 71 Subpart H regulations as to packaging must be submitted to the Department at the address provided in 24 RCNY Health Code § 175.01(c).
(f) Transport authorized by the incorporated 10 C.F.R. § 71.13 must not be by public modes of transportation including, but not limited to, buses, subways, trams, taxicabs, car services, trains, ferries, or other public conveyance, which would be returned immediately to public use after transporting licensed material.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.107. Physical Protection of Category 1 and Category 2 Quantities of Radioactive Material
(a) Except as set forth in subdivision (b) of this section, 10 C.F.R. Part 37 is hereby incorporated by reference herein to this Article with the same force and effect as if fully set forth in its entirety.
(b) The following provisions from 10 C.F.R. Part 37 are not so incorporated: § 37.1, § 37.3, § 37.7, § 37.9, § 37.11(a-b), § 37.13, § 37.105, § 37.107 and § 37.109.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 175.108. Quality Assurance Programs
(a) Purpose and scope. This section establishes requirements for the use of radioactive materials or the radiation therefrom for diagnostic and therapeutic uses in the healing arts. These requirements provide for the protection of the public health and safety and are in addition to, and not in substitution for, other requirements in this Article. The requirements of this section apply to all licensees subject to this Article.
(b) Diagnostic facilities. A quality assurance program is a system of plans, actions, reviews, reports and records, the purpose of which is to ensure that diagnostic facilities achieve consistent high quality imaging and other diagnostic results, while maintaining personnel doses within limits prescribed by the Department.
(1) Each licensee performing diagnostic procedures must implement a quality assurance program, which includes at a minimum:
(i) the adoption of a manual containing written policies and procedures for radiation protection and describing the licensee's quality assurance program that is facility- and equipment-specific. Policies and processes must be consistent with the types of procedures provided including, but not limited to, identification of patients, personnel monitoring, and protection of pregnant workers and patients. The quality assurance manual must describe the various processing, generator and systems quality control tests appropriate for the types of procedures provided in sufficient detail to ensure that they will be performed properly;
(ii) the performance of quality control tests and the correction of deficiencies as specified in the quality assurance manual;
(iii) the provisions of a formalized in-service training program for employees including, but not limited to, quality assurance and radiation safety procedures;
(iv) the measurement of the amount of activity of each dose of a radiopharmaceutical/ radiobiologic administered to each patient;
(v) the provision of the information described in 24 RCNY Health Code § 175.108(b)(1)(iv) to any patient upon request; and
(vi) the performance of an ongoing program of analysis of repeated, rejected or wrongly administered diagnostic studies which are designed to identify and correct problems and to optimize quality.
(2) Each licensee must maintain written records documenting quality assurance and audit activities for review by the Department. Unless otherwise required, such records must be maintained by the licensee until after the next scheduled inspection is completed by the Department.
(c) External beam and brachytherapy. A quality assurance program for external beam therapy or brachytherapy is a system of plans, actions, reviews, reports and records, the purpose of which is to ensure a consistent and safe fulfillment of the dose prescription to the target volume with minimal dose to normal tissue.
(1) Each licensee authorized to administer external beam therapy or brachytherapy to humans must implement a quality assurance program to systematically monitor, evaluate and document radiation therapy services to ensure consistent and safe fulfillment of the dose prescription to the target volume, with minimal dose to normal tissue, minimal exposure to personnel and adequate patient monitoring aimed at determining the end result of the treatment. Each licensee must meet or exceed all quality assurance criteria described in this subdivision.
(2) Each licensee must adopt and maintain a quality assurance program that includes policies and procedures that require the following:
(i) Each patient's medical record must be complete, accurate, legible and must include the patient's initial clinical evaluation, treatment planning data, treatment execution data, clinical assessments during treatment, a treatment summary and plan for subsequent care. Treatment related data must be recorded in the patient's medical record at the time of each treatment.
(ii) A written directive or dated order or prescription for the medical use of radiation or radioactive material must be made for each patient in accordance with 10 C.F.R. § 35.40. The directive, order or prescription must be signed or approved electronically by a board certified radiation oncologist or qualified physician who restricts their practice to radiation oncology.
(iii) The accuracy of treatment plan data and any modifications to treatment plan data transferred to a radiation treatment delivery system must be verified by qualified clinical staff prior to patient treatment.
(iv) A radiation therapy technologist, physician or other qualified health practitioner must verify that the patient set up on the treatment machine is in accordance with the treatment plan prior to the first fraction of a course of treatment and prior to treatment for any changes to the initial treatment plan.
(v) Clinical staff must obtain clarification before beginning a patient's treatment if any element of the order or other record is confusing, ambiguous, erroneous or suspected of being erroneous.
(vi) Each patient's identification must be verified by at least two different means by qualified clinical staff prior to each treatment.
(vii) Each patient's response to treatment must be assessed by a board certified radiation oncologist or other qualified physician in the active practice of external beam therapy or brachytherapy. Unusual responses must be evaluated as possible indications of treatment errors and recorded in the patient's medical record.
(viii) The medical records of patients undergoing fractionated treatment must be checked for completeness and accuracy by qualified clinical staff at intervals not to exceed six fractions.
(ix) Radiation treatment plans and related calculations must be checked by qualified clinical staff for accuracy before 25 percent of the prescribed dose for external beam therapy or 50 percent of the prescribed dose for brachytherapy is administered, except the check must be performed prior to treatment for: any single fraction treatment; any fractional dose that exceeds 300cGy or 700 monitor units; or when the output of a medical therapy accelerator exceeds 600 monitor units per minute during treatment. If a treatment plan and related calculations were originally prepared by a board certified radiation oncologist or an authorized medical physicist possessing the qualifications specified in 10 C.F.R. § 35.51, it may be rechecked by the same individual using a different calculation method. Treatment plans and related calculations prepared by other qualified clinical personnel must be checked by a second qualified person using procedures specified in the treatment planning procedures manual required pursuant to paragraph (3) of this subdivision and who has received training in use of this manual.
(x) All equipment and other technology used in planning and administering radiation therapy must function properly and safely, and must be calibrated properly and repaired and maintained in accordance with the manufacturer's instructions. The equipment and technology that is subject to such quality control, includes but is not limited to: computer software and hardware including upgrades and new releases; equipment used to perform simulation; dosimetry equipment; equipment used to guide treatment delivery, including but not limited to ultrasound units, kV and mV imaging equipment and monitors that are used to view patient imaging studies and personnel radiation safety equipment. Data communication between various systems, including but not limited to treatment planning systems, treatment delivery systems and data networks/storage media must be evaluated and tested to ensure accurate and complete data transfer.
(xi) Quality control tests performed on equipment and technology used in planning and implementing radiation treatment must be documented, including:
(A) Detailed procedures for performing each test;
(B) The frequency of each test;
(C) Acceptable results for each test;
(D) Corrective actions taken;
(E) Record keeping and reporting procedures for test results including the tester's name, signature and date of the test; and
(F) The qualifications are specified for the individual conducting the test and for the person who reviews test data.
(xii) Test results that exceed tolerances/limits must be immediately reported to the authorized medical physicist or QMP.
(xiii) Records for all maintenance, repairs and upgrades of equipment and technology must be maintained for at least 5 years.
(xiv) Errors or defects in technology or equipment, including computer hardware and software, must be reported to the technology or equipment manufacturer and to the United States Food and Drug Administration (MedWatch) as soon as possible and in no event more than 30 days of discovery, and records of equipment errors and reports required by this subparagraph must be maintained for review by the Department for at least 3 years.
(xv) Patients with permanent brachytherapy implants must be provided with instructions to take radiation safety precautions, as required by 10 C.F.R. § 35.75 and the licensee's radioactive materials license, after being released from the licensee's facility.
(xvi) All personnel involved in planning or implementing radiation therapy must be credentialed and acting within their scope of practice as provided in New York State law. Credentialing must include verifying that all professional staff are appropriately licensed, including medical physicists and radiation therapy technologists. Records of credentialing must be maintained during the period in which the credentialed person provides services to the licensee or registrant, and for 3 years thereafter.
(xvii) Any unintended deviation from the treatment plan that is identified must be evaluated and corrective action to prevent recurrence must be implemented. Records of unintended deviations and corrective action must be maintained for audits required by paragraph (5) of this subdivision and for review by the Department. (xviii) There must be a process to ensure quick and effective response to any radiation therapy related recalls, notices, safety alerts and hazards.
(3) Each licensee must adopt and maintain a radiation treatment manual prepared by an authorized medical physicist possessing the qualifications specified in 10 C.F.R. § 35.51. The manual must include the calculation methods and formulas to be used at the facility (including the methods for performing the checks of treatment plans and related calculations as required by clause (I) of subparagraph (i) of paragraph (3) of this subdivision). The treatment planning manual may be part of the quality assurance manual. The radiation treatment manual must be included in training given pursuant to 10 C.F.R. § 19.12 to facility staff who will participate in treatment planning. Each licensee must ensure that an authorized medical physicist possessing the qualifications specified in 10 C.F.R. § 35.51 prepares or reviews and approves a procedures manual describing how radiation therapy treatment planning is to be performed at the licensee's facility and reviews the treatment planning manual at least annually.
(4) Each licensee must ensure that all equipment used in planning and administering radiation therapy is functioning properly, designed for the intended purpose, properly calibrated, and maintained in accordance with the manufacturer's instructions and the quality assurance program described in the licensee's quality assurance manual. Such equipment must be calibrated prior to use on patients, at least annually thereafter and following any change, repair or replacement of any component which may alter the radiation output.
(5) Each licensee must implement written procedures for auditing the effectiveness of the radiation therapy quality assurance program that includes the following:
(i) Audits must be conducted at intervals not to exceed 12 months by an authorized medical physicist possessing the qualifications specified in 10 C.F.R. § 35.51, and also by a physician, both of whom are in the active practice of the type of radiation therapy conducted by the licensee; and
(ii) The licensee must ensure that the individuals who conduct the audit prepare and deliver to the licensee a report which contains an assessment of the effectiveness of the quality assurance program and makes recommendations for any needed modifications or improvements; and
(iii) The licensee must promptly review the audit findings, address the need for modifications or improvements and document actions taken. If recommendations are not acted on, the licensee must document the reasons therefor and also any alternative actions taken to address the audit findings; and
(iv) Each licensee must maintain complete written records relating to quality assurance and audit activities for review and inspection by the Department. Audit records must be maintained for at least 6 years.
(6) Accreditation in Radiation Oncology.
(i) Each licensee must maintain accreditation in radiation oncology by the American College of Radiology, the American College of Radiation Oncology or other equivalent accrediting organization as determined by the Department.
(ii) The licensee must maintain a record of accreditation, including a copy of the application, all supplemental application information and all correspondence transmitted between the accrediting body and the licensee. Records must be maintained for at least 6 years.
(d) Unsealed byproduct material for which a written directive is required. A quality assurance program for unsealed byproduct material for which a written directive is required is a system of plans, actions, reviews, reports and records, the purpose of which is to ensure a consistent and safe fulfillment of the dose prescription.
(1) Each licensee who uses unsealed byproduct material for which a written directive is required in humans must implement a quality assurance program which includes at a minimum:
(i) The adoption of a manual containing written policies and procedures designed to assure effective supervision, safety, proper performance of equipment, effective communication and quality control. The manual must include procedures to assure that:
(A) each patient's evaluation and intended treatment is documented in the patient's record;
(B) a written, signed and dated order for medical use of radioactive material is made in accordance with 10 C.F.R. § 35.40;
(C) each patient is positively identified;
(D) all orders and other treatment records are clear and legible;
(E) staff must be instructed to obtain clarification before treating a patient if any element of the order or other record is confusing, ambiguous or suspected of being erroneous;
(F) each patient's response to treatment must be assessed by an authorized user physician, or a physician under the supervision of an authorized user physician, for unsealed byproduct material for which a written directive is required and that unusual responses are evaluated as possible indications of treatment errors; and
(G) complete treatment records containing data recorded at the time of each treatment are maintained.
(2) Each licensee must ensure that all equipment used in planning and administering unsealed byproduct material for which a written directive is required is designed and used for the intended purpose and is properly functioning, is properly calibrated and is maintained in accordance with the manufacturer's instructions and the QA program described in the licensee's QA manual.
(3) Each licensee must audit its activities related to the use of unsealed byproduct material for which a written directive is required as part of its quality assurance program at intervals not to exceed 12 months to assess the effectiveness of the program, document the audit findings and any modifications or improvements found to be needed and institute corrective actions and improvements as indicated by the audit findings.
(Added City Record 4/24/2019, eff. 5/24/2019)
§ 177.01. Applicability
(a) The requirements of this Article apply to all tanning facilities, as defined in 24 RCNY Health Code § 177.03, including, but not limited to, those located in tanning parlors and salons, hair and nail salons, gymnasia and health establishments, apartment houses, condominiums, country clubs, or hotels.
(b) This Article does not apply to facilities where ultraviolet radiation devices are used by a qualified health care professional to treat medical conditions.
§ 177.03. Definitions
"Adequate" means sufficient to accomplish the purpose for which something is intended, and to such a degree that no reasonable risk to health or safety is presented. An item installed, maintained, designed and assembled, an activity conducted, or an act performed, in accordance with generally accepted standards, principles or practices applicable to a particular trade, business, occupation or profession, is adequate within the meaning of this Article.
"Approved" means acceptable to the New York City Department of Health and Mental Hygiene based on a determination of conformance with applicable documented standards the Department has determined, or which are set forth by any applicable regulatory agency or recognized standards-issuing body.
"C.F.R." means the Code of Federal Regulations.
"Disinfect" means adequate antimicrobial treatment by a disinfectant capable of destroying pathogenic non-spore forming bacteria, viruses, fungi, parasites and protozoa on treated surfaces.
"FDA" means the United States Food and Drug Administration.
"Formal training" means a course of instruction approved by the Department, and conducted by a person possessing adequate knowledge, training, and curriculum and certification testing experience pertaining to the correct and safe operation of an ultraviolet radiation device. A formal training course must be at least 4 hours and conclude with an exam of the information presented in the course. Successful completion of such training is dependent on passage of such exam. A course may include classroom instruction, correspondence learning or online certification.
"Operator" means a competent individual who is at least 18 years of age and who either owns a tanning facility or is designated by the owner of a facility to be responsible for operating the facility in compliance with this Article.
"Patron" means a person 17 years of age or older who uses an ultraviolet radiation device at a tanning facility.
"Permit" means a license issued to a tanning facility pursuant to this Article.
"Protective eyewear" means equipment designed to be worn by users of an ultraviolet radiation device to reduce exposure of the eyes to radiation emitted by the product. The spectral transmittance of the eyewear must meet the performance requirements of 21 C.F.R. § 1040.20(c)(4)(ii) or any successor regulation.
"Qualified health care professional" means a physician licensed by the State of New York to practice medicine, or a physician assistant or nurse practitioner licensed to practice in New York under the supervision of and/or in collaboration with a licensed physician.
"Sanitize" means adequate antimicrobial treatment by a disinfectant determined to be capable of reducing the number of pathogenic spore forming bacteria on treated surfaces. Exposure to the ultraviolet radiation produced by the ultraviolet radiation device itself is not considered an adequate sanitizing agent.
"State" means the New York State Department of Health.
"Tanning facility" means any establishment located in New York City where one or more ultraviolet radiation devices are used, offered, or made available for use by any human being, whether or not a fee is charged directly or indirectly.
"Timer" means any device incorporated into an ultraviolet radiation device that terminates radiation emission after a preset time interval.
"Ultraviolet radiation device" or "tanning device" means any product which is designed to emit electromagnetic radiation in the wavelength interval of two hundred (200) nanometers to four hundred (400) nanometers in air, and which is intended to induce tanning of the human skin through irradiation, including, but not limited to, a sunlamp, tanning booth, or tanning bed.
"Ultraviolet radiation measurement device" means any device the Department deems adequate to measure the physical characteristics of the emissions of an ultraviolet device. The device must conform to FDA-recommended standards and be calibrated according to National Institutes of Standards and Technology (NIST) recommendations.
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.05. General Requirements
(a) A tanning facility in the city cannot be in operation unless the facility has been issued a permit by the Department.
(b) A facility's permit to operate must be conspicuously posted within the tanning facility.
(c) In addition to the permit fee prescribed in Section 72-1.4(a) of Title 10 of the New York Codes, Rules and Regulations ("NYCRR"), an application for a permit must also be accompanied by payment of an inspection fee, as authorized by 10 NYCRR § 72-1.5(b), in the amount of $200 for each ultraviolet radiation device at the tanning facility.
(d) A permit issued pursuant to this Article will be issued to a specific person and will be valid only for a specified location.
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.07. Enforcement
(a) Inspections. Each operator will allow the Department to inspect the tanning facility, its equipment and records when the facility is doing business.
(b) Inspection reports. The tanning facility will maintain the inspection report provided by the Department in its records until its next inspection.
(c) Public health hazards. Where one or more of the following public health hazard conditions exists, the Department will order immediate correction, or may order the facility, or any portion of it, to immediately close. Any facility that is ordered to close may not reopen until the hazardous condition(s) that were the basis of the order are corrected to the satisfaction of the Department. Public health hazards that may result in an order to immediately close are:
(1) Wiring or electrical system components that have not been maintained, such that an imminent fire or shock hazard exists;
(2) Any ultraviolet radiation device that is not adequately labeled;
(3) Any ultraviolet radiation device that is not being operated in accordance with its label, FDA-certified manufacturer's recommendations and operating manual, or any provision of this Article;
(4) Failure to assure and maintain the accuracy of ultraviolet radiation device timers;
(5) Failure to ensure that patrons possess adequate protective eyewear;
(6) Failure to provide adequate sanitizing of tanning beds, tanning booths, pillows or headrests; or inadequate disinfection of reusable protective eyewear;
(7) Failure to provide timer lockout or remote timer controls; or
(8) Any other condition determined by the Department to be an imminent risk to the public's health and safety.
(d) Violations and penalty. In lieu of revoking, suspending or annulling a permit, the Department may assess a civil penalty of two hundred fifty dollars for any violation of this Article.
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.09. Modification by Department
When the Department determines that the strict application of any provision of this Article presents practical difficulties, the Commissioner or their designee may, in a specific instance, modify the application of such provision consistent with the general purpose of this Article and upon such condition(s) the Department deems necessary to protect life and health. Failure to comply with the terms of a modification may render such modification null and void and result in enforcement by the Department. The denial of a request for modification by the Commissioner shall be deemed a final agency determination.
(Amended City Record 6/13/2025, eff. 7/13/2025)
§ 177.11. Operator Responsibilities
(a) An operator must be present at a tanning facility whenever any ultraviolet radiation device is available for use by a patron. The operator and all employees who are authorized to operate ultraviolet radiation devices must have successfully completed formal training before operating any ultraviolet radiation device.
(b) The operator must limit patrons' exposure time as recommended by the tanning device manufacturer on the label for such device and in the operating instruction manual such that a patron may not exceed the maximum exposure time within any 24-hour period.
(c) The operator must perform annual tests on all tanning device timers to ensure that the requirements of 24 RCNY Health Code § 177.15(c)(2)(i) through (v) are met. Timer tests must be documented and recorded as required by 24 RCNY Health Code § 177.17.
(d) The operator must inform each patron of the location of the tanning device termination switch.
(e) The operator must ensure that each patron using an ultraviolet radiation device possesses adequate protective eyewear; such protective eyewear must comply with 21 C.F.R. § 1040.20(c)(4) or any successor regulation.
(f) Each tanning facility must maintain and make available to the Department upon request a list of its operators and, for each operator, a certificate of formal training showing that the operator was trained in accordance with this section.
(1) The Department will approve formal training courses that can issue certificates of formal training for operators. The Department will maintain a list of approved operator training courses on its website.
(g) If the operator maintains any web page that lists, advertises or otherwise displays the indoor tanning services available at a tanning facility, or any web page through which a prospective patron may reserve an indoor tanning service at a tanning facility, such web page must contain the following disclaimer message, which must be clearly displayed in minimum 10-size font of contrasting color:
UV radiation from indoor tanning devices can cause: • Skin Cancer, including melanoma, the type of skin cancer responsible for the most deaths • Eye burns that can cause intense pain and negatively affect vision • Sunburn (discomfort, pain and tenderness on the skin) • Early skin aging, such as wrinkles and age spots
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.13. Patron Identification and Acknowledgements
(a) Patron identification and age verification.
(1) An operator must require that every patron provide a driver's license, or other form of photo identification issued by a government entity or educational institution, indicating that the patron is at least 18 years of age.
(2) No one under eighteen (18) years of age will be permitted to use an ultraviolet radiation device in a tanning facility.
(3) The operator must conspicuously post a sign in or near the facility reception area that reads in prominent print:
PERSONS UNDER 18 YEARS OF AGE ARE PROHIBITED FROM USING UV RADIATION DEVICES. PERSONS 18 YEARS OF AGE OR OLDER MUST PROVIDE A DRIVER LICENSE OR OTHER PHOTO IDENTIFICATION ISSUED BY A GOVERNMENT OR EDUCATIONAL INSTITUTION BEFORE USING UV RADIATION DEVICES.
(b) Health Risk Advisory. During the patron's initial visit to the tanning facility, the operator must provide the Department's Health Risk Advisory to the patron. The Health Risk Advisory advises the patron of the health risks associated with the use of an ultraviolet radiation device.
(1) The Department will make available a copy of the current Health Risk Advisory with which the operator may make sufficient copies for all patrons. The copies must be the same size of the original Health Risk Advisory provided by the Department.
(c) Statement of Acknowledgement. No patron may undergo ultraviolet radiation exposure at a tanning facility without reading and signing a Statement of Acknowledgement, in a form prescribed by the Department, that meets the following requirements:
(1) The statement of acknowledgement must declare that the patron has read and understands the Health Risk Advisory.
(2) The patron agrees to wear adequate protective eyewear during the entire ultraviolet radiation exposure.
(3) The operator or a designated employee must also sign and date the statement.
(4) The statement of acknowledgement expires twelve (12) months from the date it was signed. The original signed consent form must be retained by the facility for a period of twelve (12) months and may be retained off-premises provided that an electronic image or copy of the original signed consent form is readily available to the owner, operator or employee responsible for the operation of the ultraviolet radiation device of such facility.
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.15. Facilities and Equipment
Each tanning facility must meet the following minimum requirements:
(a) Required Signs and Labels.
(1) Warning Signs. For each ultraviolet radiation device in the facility, there must be a warning sign posted in the immediate vicinity of the device. The Department will provide the warning signs to the operator. Warning signs must be: within three feet of the device, at eye level, readily legible, clearly visible, and not obstructed by any barrier, equipment, or other item present so that the patron can easily view the warning sign before energizing the ultraviolet radiation device.
(2) Ultraviolet Radiation Device Label. Each ultraviolet radiation device must have a permanent label affixed or inscribed on the exterior of the device, as required by the FDA. The device label must be clearly legible and may not be obscured, altered or tampered with.
(b) Instruction Manual. For each ultraviolet device in use in the tanning facility, a current manufacturer's operating instruction manual must be maintained onsite by the operator. The operating instruction manual must meet the applicable FDA requirements of 21 C.F.R. § 1040.20(e) or any successor regulation.
(c) All ultraviolet radiation devices must be adequately maintained and operated to meet the manufacturer's recommendations, and to meet the following minimum requirements:
(1) Label. Each ultraviolet radiation device must be adequately labeled as specified under paragraph (2) of subdivision (a) of this section.
(2) Timer. Each ultraviolet radiation device must have a timer that meets the following minimum requirements:
(i) Each ultraviolet radiation device must incorporate a timer system with multiple timer settings as specified on the manufacturer's label. The maximum timer interval(s) may not exceed the manufacturer's maximum recommended exposure time.
(ii) No timer interval may have an error greater than ± 10% of the maximum timer interval for the product.
(iii) The timer may not automatically reset and cause radiation emission to resume for a period greater than the unused portion of the timer cycle when emission from the ultraviolet lamp has been interrupted.
(iv) Only the operator or a designated employee is allowed to set the device timer.
(v) Facilities must have remote timer controls or a lock out device prior to the operation of ultraviolet radiation devices.
(3) Each ultraviolet radiation device must allow the patron using such device to manually terminate ultraviolet radiation emission at any time by using a termination switch and without disconnecting the electrical plug, removing the ultraviolet lamp or leaving the immediate environs of the ultraviolet radiation device.
(4) All ultraviolet radiation devices must be free of electrical hazards.
(5) All ultraviolet lamps must be shielded according to the manufacturer's specifications to protect patrons from injury caused by touching or breaking lamps.
(6) For stand-up booths:
(i) There must be physical barriers or other means, such as handrails or floor markings, to indicate the recommended exposure distance between ultraviolet lamps and the patron's skin.
(ii) Doors must open outwardly. Handrails and non-slip floors must be provided.
(7) The temperature within the ultraviolet device must remain below 100 degrees Fahrenheit during the operation of the device.
(8) Defective or burned out ultraviolet lamps or filters must be replaced with a type compatible for use in that device, as specified on the product label on the ultraviolet radiation device or as recommended by the manufacturer's original specifications. Replacement lamps or filters must be "compatible" as provided in 21 C.F.R. § 1040.20(e) or any successor regulation. Replacement of lamps and compatibility documentation must be recorded as a part of the maintenance log specified in 24 RCNY Health Code § 177.17.
(9) Equipment must be regularly maintained according to the manufacturer's recommendations.
(10) All ultraviolet radiation devices must meet the irradiance limitations set forth by the FDA performance requirements provided in 21 C.F.R. § 1040.20(c)(1) or any successor regulation. All ultraviolet radiation devices may not transmit measurable radiation in wavelengths less than 200 nanometers.
(11) All ultraviolet radiation devices must be maintained and operated so that the manufacturer's recommended maximum exposure time does not result in an exposure which exceeds the limits of Minimal Erythema Dose (MED) or Minimal Melanogenic Does (MMD) as set forth by the FDA and as measured by the Department with an approved measuring device and calculated according to the current FDA procedure (Food and Drug Administration Policy on Maximum Timer Interval and Exposure Schedule for Sunlamp Products, 8/21/86) or its successor.
(d) Protective eyewear.
(1) The operator must have available for patron use an adequate number of sets of protective eyewear at no additional charge to patrons. Alternatively, patrons may use their own protective eyewear.
(2) The protective eyewear that the operator provides, unless it is single-use disposable eyewear, must be cleaned and then disinfected after each use as specified in subdivision (e) of this section.
(e) Sanitation. The operator must maintain all facilities in a sanitary condition. The facilities must meet the following minimum requirements:
(1) Ultraviolet radiation devices and protective eyewear must be cleaned and then sanitized after each use, according to the following minimum provisions:
(i) The ultraviolet radiation device
A. A clean paper or cloth towel must be used each time the tanning device is cleaned and sanitized; and
B. The disinfectant must be one specifically manufactured for sanitizing ultraviolet light-emitting equipment and must be prepared and used according to manufacturer's specifications.
(ii) The protective eyewear must be cleaned with disinfectant specifically manufactured for sanitizing ultraviolet radiation protective eyewear and must be prepared and used according to the manufacturer's specifications.
(iii) Linens and other cloth.
A. Pillows and headrests must be covered in an easily cleanable material and must be cleaned and then sanitized with an adequate disinfectant after each use; and
B. If towels or other linens are provided for patron use, they must be washed with a detergent in hot water, rinsed, and thoroughly dried after each use.
(2) When the operator dilutes a concentrated disinfectant instead of using a commercially prepared, full-strength disinfectant, it must be done in accordance with the manufacturer's recommendations. A test kit or other device that accurately measures the concentration of the disinfectant in parts per million (ppm) must be used to measure the strength of the solution. The diluted disinfectant must be tested when initially prepared and at least weekly after that to ensure it continually meets the minimum concentration requirements of the manufacturer's recommendations.
(3) Written procedures maintained at the facility must include proper mixing and handling instructions for each disinfectant used to ensure proper concentration and safe use of the disinfectant.
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.17. Record Keeping
(a) Patron record. The facility must maintain a record of each patron's tanning visits, recording the date, duration of tanning exposure, and ultraviolet radiation device used and the name of person who assisted the patron in use of the ultraviolet radiation device. The facility must maintain each record on a form provided by the Department for a period of at least two (2) years after the date of the patron's last visit.
(b) The operator must keep and maintain a log of the equipment maintenance required by 24 RCNY Health Code § 177.15(c)(9). The operator must maintain the equipment log for a minimum of two (2) years and must produce such log upon Department inspection of the facility or upon Department request.
(c) The operator must maintain records showing the results of annual timer tests as detailed in 24 RCNY Health Code § 177.11(c). The operator must maintain each record for a minimum of two (2) years, and such records must be kept on site and made available to the Department immediately upon request, subject to applicable law. If such records are maintained electronically, Department staff must be allowed to access such records while on-site, subject to applicable law.
(d) The operator must maintain all records and reports required by this Article on the premises of the facility, unless an alternative is provided for in this Article, and must make them available for review by the Department on request.
(e) Records required by this section may be stored by the operator in electronic format, provided that such format can, upon inspection of the facility or request by the Department, print or produce a file in portable document format (PDF) containing the individual records required by this section.
(Amended City Record 11/4/2022, eff. 12/4/2022)
§ 177.19. Injury or Illness Incident Reporting
(a) Twenty-Four Hour Notification. The operator must report any injury or illness incidents occurring as a result of using an ultraviolet radiation device to the Department within twenty-four (24) hours of its occurrence. Reportable injuries and illnesses include, but are not limited to:
(1) all eye injuries requiring medical attention;
(2) all burns requiring medical attention;
(3) any other injury or illness incident resulting from the use of an ultraviolet radiation device for which medical care has been obtained.
(b) Report. The incident report required by subdivision (a) of this section must be in the form and manner prescribed by the Department and must include:
(1) The name of the operator;
(2) The date, time and description of the incident;
(3) The name and contact information of the affected individual;
(4) Information on the device involved in the injury, including the serial number, model number, and type of ultraviolet lamps installed in the device;
(5) The nature, cause, and extent of the alleged injury and the duration of the ultraviolet radiation exposure;
(6) The name and address of the health care provider and treatment administered, if known;
(7) Actions taken by operator or other employees at the facility; and,
(8) Any other information that may be requested by the Department.
(c) All injury and illness incident reports must be maintained at the tanning facility for a minimum of two (2) years from the date of the injury or illness and must be made available for review by the Department on request.
§ 177.21. Severability
If any clause, sentence, paragraph, subdivision, section or part of this Article shall be adjudged by any court of competent jurisdiction to be invalid, such judgment shall not affect, impair, or invalidate the remainder thereof, but shall be confined in its operation to that clause, sentence, paragraph, subdivision, section or part thereof directly involved in the controversy in which such judgment shall have been rendered.
§ 181.01. Definitions
In this article
(a) public transportation facility is used as defined in 24 RCNY Health Code § 139.01,
(b) commercial premises is used as defined in 24 RCNY Health Code § 135.01,
(c) food establishment is used as defined in 24 RCNY Health Code § 81.03,
(d) day care service, school and children's institutions are used as defined in 24 RCNY Health Code § 45.01, and
(e) hotel, rooming house and lodging house are used as defined in Multiple Dwelling Law § 4.
§ 181.03. Spitting Prohibited. [Repealed]
(Repealed City Record 9/20/2016, eff. 10/20/2016)
§ 181.05. Common Towel Prohibited
No person who owns or is in charge of commercial premises, an office or other business establishment, day care service, school, children's institution, hotel, rooming house, lodging house, public wash room, public lavatory, public transportation facility or any other public place or any place used in common by the public shall furnish or maintain or permit the furnishing or maintenance of a common towel for the use of more than one person.
§ 181.07. Common Eating and Drinking Utensils Prohibited
The use or furnishing for use of common eating or drinking utensils is prohibited in commercial premises, a public transportation facility, food establishment, hotel, rooming house, lodging house, day care service, school, children's institution, park, street or any other public place or any place used in common by the public.
§ 181.09. Manufacture and Handling of Cigars, Cigarettes and Tobacco
(a) No person engaged in the preparation, manufacture, sorting or handling of cigars, cigarettes or tobacco intended for sale shall at any time:
(1) Touch such cigar, cigarette or tobacco with his lips, teeth or tongue; or,
(2) Moisten such cigar, cigarette or tobacco with saliva, directly or indirectly, by spitting or by use of the fingers or utensils or accessories of any kind; or,
(3) Spray or moisten such cigar, cigarette or tobacco with water or any other liquid emitted from the mouth; or,
(4) Permit such cigar, cigarette or tobacco to touch or be introduced into the nose of any person.
(b) A person engaged in the preparation, manufacture, sorting or handling of cigars, cigarettes or tobacco intended for sale shall thoroughly wash his hands with soap and warm water before beginning work, immediately after each visit to the toilet, and at all other times when necessary during the course of the work.
(c) A copy of this section shall be posted conspicuously in every place where cigars, cigarettes or tobacco are prepared, manufactured, sorted or handled.
§ 181.11. Growing of Poison Ivy and Ragweed Prohibited
No person who owns, occupies or is in charge of a lot or premises shall cause or allow poison ivy, ragweed, or other poisonous or allergenic weed to grow on such lot or premises.
§ 181.15. Tattooing Prohibited. [Repealed]
BetaNYC's copy has no text for this section. Read 24 RCNY Health Code § 181.15 at American Legal Publishing.
§ 181.17. Smoking Prohibited in Certain Areas
(a) It shall be unlawful for any person to smoke or carry a lighted cigar, cigarette or pipe in any elevator or in any retail food establishment commonly known as a supermarket.
(b) It shall be unlawful for any person to smoke or carry a lighted cigar, cigarette or pipe in any classroom or in any lecture hall, except that the owner or person in charge of such classroom or lecture hall may designate a special area or areas where smoking is permitted, unless otherwise prohibited by the Fire Department or by other law. The area or areas where smoking is permitted shall contain not more than 20% of the total seats of the classroom or lecture hall.
(c) It shall be unlawful, except as provided in subsection (e), for any person to smoke or carry a lighted cigar, cigarette or pipe in any theatre, motion picture theatre, opera house, concert hall, hospital, sanatorium, nursing home, convalescent home, home for the aged or chronically ill patients, museum or library.
(d) It shall be unlawful, except as provided in subsection (e), to smoke or carry a lighted cigar, cigarette or pipe in any enclosed public space in which 50 or more persons gather for religious, recreational, political or social purpose. This subsection shall not apply to:
(1) Any place in which social functions such as weddings, parties, testimonial dinners and similar functions are held and in which the seating arrangements are under the control of the sponsor of the function and not of the owner or person in charge of such place.
(2) Any establishment which sells admission tickets on a seasonal or other periodic basis.
(e) The owner or person in charge of any building, structure or place specified in subsections (c) and (d) may designate special areas therein where smoking is permitted unless otherwise prohibited by the Fire Department or by other law.
(f) Signs prohibiting or permitting smoking, as the case may be, shall be posted conspicuously by the owner or person in charge of each building, structure or place specified in subsections (a), (b), (c) and (d).
§ 181.19. Required Point-of-Sale Tobacco Health Warnings and Smoking Cessation Information
(a) Any person in the business of selling tobacco products face-to-face to consumers in New York City shall prominently display tobacco health warning and smoking cessation signage produced by the Department.
(b) The signage required by subdivision (a) of this section shall be:
1. designed by the Department and may include:
a. information about tobacco products and the adverse health effects of tobacco use;
b. a pictorial image illustrating the effects of tobacco use; and
c. information about how to get help to quit using tobacco;
2. produced in two sizes:
a. one "small sign," not to exceed 144 square inches; and
b. one "large sign," not to exceed 576 square inches;
3. distributed by the Department or by the Department of Consumer Affairs, although additional signage may be ordered by calling 311.
(c) Persons who engage in face-to-face sales of tobacco products to consumers in New York City shall prominently display the signs required by subdivision (a) of this section by posting:
1. one "small sign" on or within 3 inches of each cash register or each place where payment may be made so that the sign(s) are unobstructed in their entirety and can be read easily by each consumer making a purchase; or
2. one "large sign" at each location where tobacco products are displayed so that:
a. the sign(s) are unobstructed in their entirety and can be read easily by each person considering a tobacco product purchase; and
b. in such a way that the distance between the bottom of the sign(s) and the floor shall be no less than four feet, and the distance between the top of such sign(s) and the floor shall be no more than seven feet.
(d) The Commissioner shall have the authority to modify periodically the text, images and content of signage produced by the Department pursuant to subdivision (a) of this section based on the Commissioner's determination that such modifications may contribute to the prevention or reduction of tobacco use and its harms, the prevention of tobacco sales to minors, or the correction of misinformation among consumers about the health effects of tobacco use. Persons required to post health warning and smoking cessation signage pursuant to subdivision (a) of this section shall display any modified signage upon receipt, as set forth in subdivision (c) of this section.
(e) The provisions of this section may be enforced by any authorized agent or employee of either the Department or the Department of Consumer Affairs, or successor agency. Any violation of this section issued by the Department of Consumer Affairs may be adjudicated at any tribunal authorized to hear such agency's violations.
(f) If any provision of this section, or its application to any person or circumstance, is held invalid by any court of competent jurisdiction, the remaining provisions or the application of the section to other persons or circumstances shall not be affected.
§ 181.21. Consent for Direct Oral Suction as Part of a Circumcision. [Repealed]
(Repealed City Record 9/16/2015, eff. 10/16/2016)
§ 201.01. Definitions
When used in this title:
(a) "Live birth" or "birth" means the complete expulsion or extraction from its mother of a product of conception, regardless of the duration of pregnancy, which after expulsion or extraction shows evidence of life, such as breathing, beating of the heart, pulsation of the umbilical cord or definite movement of voluntary muscles, whether or not the umbilical cord has been cut or the placenta is attached.
(b) "Person in charge of a hospital" means the officer or employee who is responsible for the administration of a hospital or similar institution and includes but is not limited to a person holding the title of chief executive officer, administrator, superintendent, director or executive director.
(c) "Hospital" means a facility or institution licensed pursuant to Article 28 of the State Public Health Law and defined as such in § 2801 of said law.
§ 201.03. Reporting Births
(a) When a live birth occurs in the City, it shall be reported to the Department as follows:
(1) If the birth occurs in a hospital or route thereto, by the person in charge of such hospital or his or her designee; or,
(2) If the birth occurs elsewhere than in a hospital or en route thereto, by the physician, licensed midwife or registered physician assistant, in attendance at or after such birth; or,
(3) If a physician, licensed midwife or registered physician assistant attends at or after the birth elsewhere than in a hospital, or en route thereto, as an associate of a hospital, by the person in charge of the hospital with which he or she is associated or by the designee of such person in charge; or,
(4) If the birth occurs without the attendance of a physician, licensed midwife or registered physician assistant, by either of the parents of the child or, if no parent is alive, by the next of kin of the child or any person present at the birth.
(b) A person required to report a live birth pursuant to paragraphs (1), (2) or (3) of subdivision (a) of this section shall file a certificate of birth and a confidential medical report, and a person required to report pursuant to paragraph (4) of subdivision (a) of this section shall file a certificate of birth only. Reports shall be filed within 5 business days after the birth with the office maintained and designated by the Department for such purposes.
(c) The person required to report a birth shall provide to the Department information that was required to be reported, but that was not so reported, within five business days of that person receiving the information.
(d) Upon a request by the Department for additional information that may be necessary to complete, clarify or verify the information required to be reported, the person required to report a birth shall provide such information to the Department within five business days of the request.
§ 201.05. Preparation and Certification of Certificate of Birth and Confidential Medical Report of Birth
(a) The certificate of birth and confidential medical report shall be prepared and certified by the person required to file the same pursuant to 24 RCNY Health Code § 201.03, but when the birth occurs in a hospital or en route thereto, the certificate and the confidential medical report shall be prepared and certified by the physician, licensed midwife or registered physician assistant in attendance or assisting, or by a certified nurse practitioner or registered professional nurse present at or after the birth, or by a designee of the person in charge of the hospital who is trained or approved by the Department. When a physician, licensed midwife or registered physician assistant attends at or after a birth elsewhere than in a hospital or en route thereto, he or she shall prepare and certify the certificate and confidential medical report. A person certifying a certificate and confidential medical report shall examine said documents for correctness of the information contained thereon and make any necessary changes.
(b) The certificate and confidential medical report shall be prepared on forms prescribed by the Board and furnished by the Department and shall contain no statement indicating the marital name or status of the mother or whether the child was born in or out of wedlock. The person preparing the certificate shall enter all information required by the form provided by the Department, except that in case of a child born out of wedlock such person shall not enter the name of the putative father unless there is submitted to the preparer a verified written consent of the putative father pursuant to § 17-166(d) of the Administrative Code or a voluntary acknowledgement of paternity pursuant to § 4135-b of the State Public Health Law. When the birth occurs in a hospital or en route thereto, the information required by the forms shall be taken from the hospital records of the case. In a hospital, worksheets provided or otherwise approved by the Department shall be used in the preparation of the certificate and confidential medical report, and if such worksheets are used by individuals other than a physician, licensed midwife, registered professional nurse, certified nurse practitioner or registered physician assistant, then such individuals shall be trained or approved by the Department. Worksheets shall be retained by the hospital for a period of at least three years from the date of the birth, and shall, upon request, be made available to the Department for inspection.
(c) All live births occurring in the City at facilities reporting 100 or more live births per year shall be reported to the Department electronically by means of computer programs specified and provided or otherwise authorized for use by the Department. All facilities at which fewer than 100 live births are reported per year may, at their election and upon approval by the Department, implement an electronic birth certificate reporting system or continue to report births on approved paper forms.
(d) All facilities required to file birth certificates electronically and facilities reporting fewer than 100 births per year which elect to report electronically, shall apply to the Department prior to implementing any system and, upon approval by the Department, shall make electronic reports of births and, on and after January 1, 2014, acknowledgments of paternity, only in such manner and on computer programs prescribed and provided or otherwise authorized by the Department. Facilities subject to this requirement must retain acknowledgments of paternity reported electronically for a period of at least three years from the date of birth and must make these records available to the Department for inspection upon request.
§ 201.07. Confidential Medical Report of Birth; Not Subject to Compelled Disclosure or Inspection
(a) The confidential medical report of birth shall be confidential and not subject to compelled disclosure or to inspection by persons other than the Commissioner or authorized personnel of the Department, except in a criminal action or criminal proceeding, or for official purposes by a federal, state, county or municipal agency charged by law with the duty of detecting or prosecuting crime. The Commissioner or the Commissioner's designee may, however, approve the inspection of such medical reports for scientific purposes.
(b) Within the context of this section, scientific purposes shall mean epidemiologic surveillance and investigation by a governmental public health agency, research, or the compilation of statistics relating factors bearing on disease incidence, prevalence, mortality or treatment.
(Amended City Record 3/25/2016, eff. 4/24/2016)
§ 201.09. Foundlings
(a) The report of the finding of a child whose parents are unknown, filed by the Commissioner of the City Administration for Children's Services in accordance with the provisions of subdivision two of § 398 of the State Social Services Law, shall constitute the birth record of such child.
(b) The address or location where such child was found shall be considered as the place of birth, and the date of birth shall be that determined by the Commissioner of the City Administration for Children's Services as the approximate date of birth.
(c) If, however, such child is subsequently identified, and a certificate of birth for such child has been duly filed either before or following identification, the report of the Commissioner of the City Administration for Children's Services shall be placed under seal by the Department, and such seal shall not be broken except upon order of a court of competent jurisdiction.
§ 201.11. Delayed Registration of Births
(a) When a birth in the City is not recorded in the Department within one year following the birth, it may be recorded with the approval of the Commissioner or the Commissioner's designee. Application for such delayed registration shall be made on a form furnished by the Department by the parents or surviving parent, or by the guardian of the person whose birth is to be recorded, if such person is a minor, or by the person himself or herself if he or she is 18 years of age or over and his or her parents are dead. The application shall be accompanied by the following:
(1) A certified statement issued by the Department that a search was made for the record of birth in question and that such record was not found; and
(2) Such documentary and other evidence as will establish to the satisfaction of the Commissioner or the Commissioner's designee the facts and date of birth as alleged in the application. The burden of submitting convincing proof rests with the applicant.
(b) When an application for delayed registration has been granted and a certificate of birth on a delayed registration form is filed pursuant to this section, the Department shall issue to the applicant without further charge, in exchange for the certified statement submitted pursuant to paragraph (1) of subdivision (a) of this section, a certified copy of the certificate of birth.
(c) No application for delayed registration shall be granted, and no delayed certificate of birth shall be registered or issued for a deceased person.
§ 203.01. Definitions
When used in this title:
(a) "Termination of pregnancy" means the expulsion or extraction of a conceptus, regardless of the duration of pregnancy, other than a live birth as defined in 24 RCNY Health Code § 201.01(a) and includes “pregnancy loss” as defined in New York State Public Health Law § 4160.
(b) "Spontaneous termination of pregnancy" means the unplanned termination of a pregnancy, including but not limited to an ectopic pregnancy, miscarriage, still birth, or such a termination associated with a cesarean section, or an operative procedure unrelated to pregnancy resulting in an inadvertent termination.
(c) "Conceptus" means the product of any termination of pregnancy, regardless of its duration, including a hydatidiform mole, fetal tissue or other evidence of pregnancy recovered by operative or other procedure, but not including a live birth as defined in 24 RCNY Health Code § 201.01(a).
(d) "Licensed health care practitioner" means a physician or other person licensed or authorized pursuant to the New York State Education Law, or other applicable law, to perform terminations of pregnancy.
(Amended City Record 12/29/2023, eff. 1/28/2024; amended City Record 5/27/2026, eff. 6/26/2026)
§ 203.03. Reporting Terminations of Pregnancy
(a) When a spontaneous termination of pregnancy occurs in the City it shall be reported as follows:
(1) If in a hospital or en route thereto, by the person in charge of such hospital or his or her designee; or
(2) If elsewhere than in a hospital or en route thereto, by the licensed health care practitioner in attendance at or after such event or by their designee; or
(3) If a licensed health care practitioner attends at or after the event elsewhere than in a hospital or en route thereto as an associate of a hospital, by the person in charge of the hospital with which the licensed health care practitioner is associated or by the designee of such person in charge; or
(4) If investigated by the office of chief medical examiner, by a medical examiner within that office.
(b) The person required to report a spontaneous termination of pregnancy pursuant to subdivision (a) of this section shall file a certificate of spontaneous termination of pregnancy, including a confidential medical report, for a spontaneous termination of pregnancy; provided that a medical examiner, when required to report pursuant to paragraph (a)(4) of this section, shall file a certificate of spontaneous termination of pregnancy only.
(c) A certificate of spontaneous termination of pregnancy required by this section shall be filed within 24 hours after the event if a disposition permit issued pursuant to Article 205 of this Code is required or requested, and in all other cases a certificate of spontaneous termination of pregnancy shall be filed with any office maintained and designated by the Department for such purposes within five business days, including within five business days after the date the medication for a medication-induced termination of pregnancy was prescribed or ordered.
(d) (1) In circumstances where the issuance of a disposition permit pursuant to 24 RCNY Health Code Article 205 is requested and a person required to report a spontaneous termination of pregnancy pursuant to subdivision (a) of this section has not filed a report thereof electronically, the requirement of filing a certificate and confidential medical report, if any, required by this section may be fulfilled by delivery of the same immediately upon demand and within the time prescribed by subdivision (c) of this section to a funeral director or undertaker authorized to take charge of the conceptus or to the person in charge of the City mortuary if the remains are to be buried in the City cemetery. Such funeral director, undertaker or person in charge of the City mortuary, or an agent of such funeral director or undertaker registered with the Department pursuant to 24 RCNY Health Code Article 205 or a designee of the person in charge of the mortuary, shall then file the certificate within 48 hours following the receipt of the certificate of spontaneous termination of pregnancy.
(2) In all other circumstances where the issuance of a disposition permit pursuant to 24 RCNY Health Code Article 205 is requested for a conceptus, a funeral director or undertaker authorized to take charge of the conceptus, or the person in charge of the City mortuary if the remains are to be buried in the City cemetery, registered with the Department pursuant to 24 RCNY Health Code Article 205 or a designee of the person in charge of the mortuary, must report within 24 hours of the termination of pregnancy to the Department only the information required on the forms prescribed by the Board and furnished by the Department.
(3) Funeral directors, undertakers, City mortuary personnel, and their agents or designees, must not divulge information in the confidential documents related to disposition of a conceptus except to authorized personnel of the Department.
(e) All spontaneous terminations of pregnancy occurring at or en route to hospitals or other facilities that report births electronically to the Department pursuant to 24 RCNY Health Code Article 201 and all spontaneous terminations of pregnancy reported by the office of chief medical examiner, shall be reported to the Department electronically by means of computer programs specified and provided or otherwise authorized for use by the Department. In circumstances where the issuance of a disposition permit pursuant to 24 RCNY Health Code Article 205 is requested, and a person required to report a spontaneous termination of pregnancy pursuant to subdivision (a) of this section files a report thereof electronically, a funeral director or undertaker authorized to take charge of the remains, or the person in charge of the City mortuary when said mortuary files an application for a disposition permit, shall also file, within 72 hours following the termination of pregnancy, the application for such a permit electronically by means of computer programs specified and provided or otherwise authorized for use by the Department. All hospitals or other facilities that are not required to report spontaneous terminations of pregnancy electronically pursuant to this subdivision may, at their election and upon approval by the Department, implement an electronic reporting system, or continue to report spontaneous terminations of pregnancy on approved paper forms. However, once a hospital or facility has commenced reporting electronically, such hospital or facility may not report on paper forms unless otherwise authorized by the Department.
(f) All facilities required or electing to report electronically pursuant to subdivision (e) of this section shall apply to the Department prior to implementing any electronic reporting system and, upon approval by the Department, shall make electronic reports only in such manner and on computer programs prescribed and provided by or otherwise authorized by the Department. Written paper reports may be submitted for a limited period of time only in the case of extenuating circumstances, temporary equipment failure, or prolonged inability to access the electronic reporting system, and only with the specific approval of the Department. In addition, the Department may, on its own initiative, allow written, paper reports to be submitted if electronic reporting is not possible in a particular circumstance, as a result of a deficiency in the Department's electronic reporting system. The Department may, in addition, require summary, cumulative or periodic reports on such reporting schedule as it may deem necessary.
(g) The person required to report a spontaneous termination of pregnancy or to file an application for a disposition permit shall provide to the Department information that was required to be reported, but that was not so reported, within five business days of that person receiving the information.
(h) Upon a request by the Department for additional information that may be necessary to complete, clarify or verify the information required to be reported, the person required to report a spontaneous termination of pregnancy or to file an application for a disposition permit shall provide such information to the Department within five business days of the request.
(Amended City Record 12/29/2023, eff. 1/28/2024; amended City Record 10/8/2024, eff. 11/7/2024; amended City Record 5/27/2026, eff. 6/26/2026)
§ 203.05. Preparation and Certification of Certificates
(a) (1) Preparation. Any certificate or confidential medical report required by this Article shall be prepared by the same person required to file the same pursuant to 24 RCNY Health Code § 203.03 but when a spontaneous termination of pregnancy occurs in a hospital or en route thereto, the certificate and confidential medical report, if any, shall be prepared by a licensed health care practitioner in attendance, assisting or present at or after the event, by the chief medical officer of the hospital, by the physician in charge of the service on which the pregnant person was treated, or by a designee of the person in charge of the hospital who is trained or approved by the Department. When a licensed health care practitioner attends at or after a spontaneous termination of pregnancy elsewhere than in a hospital or en route thereto, he or she, or a designee of such person who is trained or approved by the Department, shall prepare the required certificate and confidential medical report, if any.
(2) Certification. A certificate of spontaneous termination of pregnancy and the confidential medical report shall be certified by a physician in attendance or assisting at or after the event, by the chief medical officer of the hospital where the event occurred, or by the physician in charge of the service on which the pregnant person was treated, or by a designee of such person who is trained or approved by the Department. When a spontaneous termination of pregnancy certificate is filed by the office of chief medical examiner, the certificate shall be certified by a medical examiner within that office. A person certifying a certificate and confidential medical report, if any, shall examine said documents for correctness of the information contained thereon and make necessary changes.
(b) The certificates specified in 24 RCNY Health Code § 203.03(b), except for certificates filed electronically pursuant to 24 RCNY Health Code § 203.03(e), shall be prepared on forms prescribed by the Board and furnished by the Department. Computer programs specified and provided or otherwise authorized for use by the Department for electronic filing shall be reflective of the forms prescribed by the Board except to the extent that differences may be necessary or warranted in order to accommodate electronic formatting. The person preparing the certificate shall enter all information required by the appropriate form. When a termination of pregnancy occurs in a hospital or en route thereto, the information shall be taken from the hospital record of the case. If worksheets are used to prepare certificates of spontaneous termination of pregnancy and confidential medical reports, if any, the worksheets shall be ones provided by the Department or in a form approved by the Department. If individuals other than a physician, licensed midwife, registered professional nurse, certified nurse practitioner or registered physician assistant use such worksheets, then such individuals shall be trained or approved by the Department. The person preparing the certificate and confidential medical report, if any, or such person's employer, shall retain such worksheets for a period of three years from the date of the event, and shall, upon request, make such worksheets available to the Department for inspection.
(Amended City Record 10/8/2024, eff. 11/7/2024; amended City Record 5/27/2026, eff. 6/26/2026)
§ 203.07. Confidential Medical Report of Spontaneous Termination of Pregnancy and Certificate of Induced Termination of Pregnancy; Not Subject to Compelled Disclosure or Inspection
(a) The certificate and confidential medical report of a spontaneous termination of pregnancy and the documents submitted pursuant to 24 RCNY Health Code § 203.03(d)(2) shall be confidential and not subject to disclosure or to inspection by persons other than the Commissioner or authorized personnel of the Department. The Commissioner or the Commissioner's designee may, however, approve the inspection by others of such medical reports and certificates for scientific purposes or in accordance with federal, New York State, or New York City law.
(b) Within the context of this section, scientific purposes shall mean epidemiologic surveillance and investigation by a governmental public health agency, research, or the compilation of statistics relating factors bearing on disease incidence, prevalence, mortality or treatment, provided that, in no instance shall personally identifiable information be released.
(Amended City Record 3/25/2016, eff. 4/24/2016; amended City Record 6/22/2023, eff. 7/22/2023; amended City Record 12/29/2023, eff. 1/28/2024; amended City Record 5/27/2026, eff. 6/26/2026)
§ 203.09. Disposal of Conceptus
Upon request a conceptus may be disposed of in a manner provided for human remains generally and in accordance with a disposition permit issued pursuant to 24 RCNY Health Code Article 205.
(Amended City Record 10/8/2024, eff. 11/7/2024)
§ 205.01. Definitions
When used in this title:
(a) "Burial" means interment of human remains in the ground or in a tomb, vault, crypt, cell or mausoleum, and includes any other usual means of final disposal of human remains other than cremation.
(b) "Cremation" means the burning of human remains in a crematory.
(c) “Disposition” means the care, disposal, transportation, burial, cremation, natural organic reduction, or embalming of the body of a deceased person, and associated measures.
(d) "Human remains" or "remains" means all or any part of the dead body of a human being but does not include human ashes recovered after cremation. For purposes of this Article a conceptus of any gestational age for which the issuance of a disposition permit pursuant to 24 RCNY Health Code § 205.21 has been requested shall be treated as human remains.
(e) "Hospice" means a program issued a certificate of approval pursuant to Article 40 of the Public Health Law and defined as such in § 4002 of said law.
(f) “Interim disposition” means the non-final, temporary location of a decedent before the final disposition is determined.
(g) "Person in control of disposition" means a person who has the right to control the disposition of the remains of a decedent in accordance with the provisions of § 4201 of the Public Health Law.
(Amended City Record 10/8/2024, eff. 11/7/2024; amended City Record 9/29/25, eff. 10/29/26)